3 | 

P a g e

 

 

(12)

 

It is not allowed to dismantle or modify the Warmer. Opening the shell may cause 

 

electrocution. Maintenance or upgrade to the Warmer must be done by authorized 

 

staff from Manufacturer. 

(13)

 

Check the Warmer daily and check all the functions to determine it is in good 

 

condition if it is not used for a long period. 

(14)

 

Turn off the Warmer and contact Jorgensen Laboratories 800-525-5614 when 

 

a malfunction has occured. 

 

Introduction

 

 

The use of room temperature (70ºF/21ºC) IV fluids in a patient suffering from shock, 
or susceptibility to hypothermia should be avoided.  Ideally fluids should be 
administered as close to body temperature as possible for better therapeutic results. 
Blood products are refrigerated or frozen prior to administration. 
 

Application 
 

It is intended to provide continuous heating during the infusion and blood transfusion. 
 

Working Principle 
 

The Jorvet IV Fluid and Blood Transfusion Warmer is an indirect heating device (dry 
warmer).  The IV fluids are heated as the fluids pass through the channel of the 
warmer.  The temperature can be set between 32⁰C ~ 42⁰C (89.6⁰F ~ 107.6⁰F.  The 
heating control board will work upon the set temperature. 
 
 

Components and Function 

 
The Jorvet IV Fluid and Blood Transfusion Warmer is made up of the following 
parts: 

 

Main control system: This system controls the whole unit by managing and 
processing the system 

 

Heating device: The heating device is made up of a heating sheet and heating 
exchanger; this allow the heat to move through the exchanger to heat the fluid 

 

Detection sensor: This includes a temperature detection sensor  

 

Alarm System: The alarm includes an audio and visual alarm 

 

Input and Display: This allows for the input of the heating parameter, 
temperature, alarm message and other parameters. 

Summary of Contents for JORVET IV

Page 1: ...0 P a g e JORVET IV FLUID BLOOD INFUSION WARMER Operation Manual JORGENSEN LABORATORIES PHONE 800 525 5614 FAX 970 663 5042 info jorvet com...

Page 2: ...splayed appears either in Fahrenheit or Celsius Overheating Temperature protection Table of Contents General Information 2 Precautions 2 Warnings 2 3 Introduction 3 Application 3 Working Principle 3 C...

Page 3: ...and to prevent damage to the unit by sliding down or dropping to the floor 3 The unit should only be used with the power cord that came with the unit and is not shared during the use of the Warmer 4 I...

Page 4: ...d to provide continuous heating during the infusion and blood transfusion Working Principle The Jorvet IV Fluid and Blood Transfusion Warmer is an indirect heating device dry warmer The IV fluids are...

Page 5: ...nical Specifications Environment The Warmer must have a 5cm circumference around the unit to allow the Warmer to have heat loss and provide air recycle If the Warmer is being moved to another location...

Page 6: ...tandard single PVC IV Set 20 and 60 d ml with an outer diameter of 3 4 4 5mm Temperature Range 95 107 6 F 35 C 42 C Temperature Accuracy 1 C or 1 8 F Warming up time 2 minutes Overheat Protection 48 C...

Page 7: ...emperature alarm 6 Temperature Unit convert C and F convert 7 Mute Button Silence for 2 minutes by pressing the mute button after alarm 8 Minus Button Decrease the temperature in increments of 0 1 C 9...

Page 8: ...be placed slightly under the IV line s exit from the infusion pump 2 Use the knob to tighten to the IV Stand 3 Prime the IV set line removing all air bubbles 4 Open the door of the warmer 5 Install th...

Page 9: ...erature In the standby mode press to set the temperature in increments of 0 1 C 9 Adjusting the Temperature unit setting In the standby mode press to set C or F unit It will display once the unit is s...

Page 10: ...overheat alarm light will keep flashing Press stop the warmer and restart it If this issue continues contact Jorgensen Laboratories System Error Alarm When E001 is displayed on the LED this indicates...

Page 11: ...ck the patient 2 Confirm the cause of the alarm 3 Turn off the unit 4 Try again Alarm and Error Alarm Possible Cause Solution Displays the room temperature after switching on the warmer Door not close...

Page 12: ...functionable Alarms are functioning appropriately Accuracy with the Temperature in the standard range Please contact Jorgensen Laboratories if any are abnormal Clean and Sterilize Use a low concentrat...

Page 13: ...ic interference Do not place other equipment next to the Warmer or overlap equipment with the warmer If it is required to do so monitor the Warmer to ensure that it is working properly The Warmer must...