Alarm System Setting
Press
mute button to silence the alarm for 2 minutes, but it will not stop clear
the alarm.
Alarm Signal
The red flashing light is the alarm and can be seen up to 13 feet of the Warmer.
The audible alarm has an intensity of 45 decibels from 3 feet from the warmer.
Action for the Alarm
When the alarm sounds, do the following steps:
1.
Check the patient
2.
Confirm the cause of the alarm
3.
Turn off the unit
4.
Try again
Alarm and Error
Alarm
Possible Cause
Solution
Displays the room
temperature after switching
on the warmer
Door not closed
Close the door
Did not press Start/Stop Button Press Star/Stop Button to Start
Detection sensor or heating
system error
Call Support if the problem
does not resolve
Overheat Alarm
Direct Sunlight or infrared light
leads to overheat temperature
Avoid the sunlight or infrared
environment
Electromagnetic interference
Avoid Electromagnetic
interference environment