General Information 

The 

Jorvet

 

IV Fluid and Blood Transfusion Warmer

 is a dry- heat device.  This unit is 

designed to warm IV fluids safely through the heating plate.  The 

Warmer

 does not 

control the fluid flow rate.  Fluids can be warmed to temperatures between 95º-107.6º F 
(35ºC-42ºC) at flow rates 25ml/hr to 1000ml/hr.  
 

Precautions 

Save this Manual for Future Reference 

 

IMPORTANT: Before using the 

Jorvet

 

IV Fluid and Blood Transfusion Warmer

please read and understand this User’s Guide and the SAFETY PRECAUTIONS 
prior to use.  If you have any questions, please contact Jorgensen Laboratories at 1-
800-525-5614. 

 

Carry out the inspections and maintenance periodically according to the manual to 
avoid patient or user injury, 

 

This operation manual is only used by professional trained Veterinary Personnel. 

Warnings 

(1)

 

 

This device is used for heating the infusion, blood transfusion and can be used 

 

only by the professional trained Veterinary Personnel. 

(2)

 

 

Securely attach the Warmer to an IV Stand to prevent damage to the unit by 

 

sliding down or dropping to the floor. 

(3)

 

 

The unit should only be used with the power cord that came with the unit and is 

 

not shared during the use of the Warmer. 

(4)

 

 

It is not allowed to use voltage other than that specified on the product label, or it 

 

could cause damage or a fire. 

(5)

 

 

Inspect the Warmer and all accessories completely before use to verify it is 

 

working in a safe manner. 

(6)

 

 

Do Not only rely on the Warmer’s Alarm to notify of a problem, the Warmer 

 

should be monitored throughout the infusion. 

(7)

 

 

Stop the use of the Warmer during an alarm 

(8)

 

 

Do Not connect other brand heating devices or other brand heating accessories to 

 

this Warmer as it could lead to wrong heating temperature or dangerous infusion. 

(9)

 

 

Keep the infusion line straight and unfolded to prevent a long-term occlusion and 

 

heating to fluids/blood for long period of time. 

(10)

 

Do Not use the Warmer where there is strong sunlight, cold & hot, wind or a 

 

dusty environment. 

(11)

 

Microwave will influence the function of the Warmer. All other devices next to 

 

the Warmer must meet the EMC standard. Mobile phone, X-Ray or MRI will be 

 

the cause of a malfunction due to its high frequency radiation. 

Summary of Contents for JORVET IV

Page 1: ...0 P a g e JORVET IV FLUID BLOOD INFUSION WARMER Operation Manual JORGENSEN LABORATORIES PHONE 800 525 5614 FAX 970 663 5042 info jorvet com...

Page 2: ...splayed appears either in Fahrenheit or Celsius Overheating Temperature protection Table of Contents General Information 2 Precautions 2 Warnings 2 3 Introduction 3 Application 3 Working Principle 3 C...

Page 3: ...and to prevent damage to the unit by sliding down or dropping to the floor 3 The unit should only be used with the power cord that came with the unit and is not shared during the use of the Warmer 4 I...

Page 4: ...d to provide continuous heating during the infusion and blood transfusion Working Principle The Jorvet IV Fluid and Blood Transfusion Warmer is an indirect heating device dry warmer The IV fluids are...

Page 5: ...nical Specifications Environment The Warmer must have a 5cm circumference around the unit to allow the Warmer to have heat loss and provide air recycle If the Warmer is being moved to another location...

Page 6: ...tandard single PVC IV Set 20 and 60 d ml with an outer diameter of 3 4 4 5mm Temperature Range 95 107 6 F 35 C 42 C Temperature Accuracy 1 C or 1 8 F Warming up time 2 minutes Overheat Protection 48 C...

Page 7: ...emperature alarm 6 Temperature Unit convert C and F convert 7 Mute Button Silence for 2 minutes by pressing the mute button after alarm 8 Minus Button Decrease the temperature in increments of 0 1 C 9...

Page 8: ...be placed slightly under the IV line s exit from the infusion pump 2 Use the knob to tighten to the IV Stand 3 Prime the IV set line removing all air bubbles 4 Open the door of the warmer 5 Install th...

Page 9: ...erature In the standby mode press to set the temperature in increments of 0 1 C 9 Adjusting the Temperature unit setting In the standby mode press to set C or F unit It will display once the unit is s...

Page 10: ...overheat alarm light will keep flashing Press stop the warmer and restart it If this issue continues contact Jorgensen Laboratories System Error Alarm When E001 is displayed on the LED this indicates...

Page 11: ...ck the patient 2 Confirm the cause of the alarm 3 Turn off the unit 4 Try again Alarm and Error Alarm Possible Cause Solution Displays the room temperature after switching on the warmer Door not close...

Page 12: ...functionable Alarms are functioning appropriately Accuracy with the Temperature in the standard range Please contact Jorgensen Laboratories if any are abnormal Clean and Sterilize Use a low concentrat...

Page 13: ...ic interference Do not place other equipment next to the Warmer or overlap equipment with the warmer If it is required to do so monitor the Warmer to ensure that it is working properly The Warmer must...