background image

4    RLD-H10G Product/Technical Bulletin

C

alibration and Verification

The leak vial may be used as often as necessary to
assure proper operation of the RLD-H10G, and to
assist in determining the size of the leak.  When using
the small leak range position, a leak of HFC causing
an audible signal equal to that produced by the
leak vial has a leak rate of approximately
one-half ounce/year.  When using the medium leak
range position on a CFC or HCFC system, this leak
rate would also be approximately one-half ounce/year.

Proper unit response during the calibration test verifies
sensitivity and operation for all three test positions.

1.

Ensure that the range switch is in the small (“SM”)
position and idling at three or four beats per
second.  (See Figure 8.)

2.

Remove the black, sticky-label disc from the cap of
the leak vial and discard.  Allow a few minutes for
the leak rate to stabilize.  Do not attempt to
unscrew the cap, as you may damage the bottle.
The vial contains enough refrigerant to last for
approximately six months.

3.

Test operation by briefly moving the probe tip over
the opening of the leak vial as if you are trying to
locate this leak.  It is not necessary to hold the
probe over the leak, as this does not simulate field
conditions.

The RLD-H10G should respond with an increase
in flash rate and sound, verifying operation and
indicating the location of the leak.

4.

If the RLD-H10G does not respond to the leak vial,
make sure there is liquid in the bottle.  If the unit
still does not respond to the leak, increase the
sensor heat slightly by turning the sensor heat
control knob (on the bottom of the unit) 1/16th of
a turn clockwise.

5.

Allow a minute for the sensor to heat, then repeat
the calibration procedure.

T

ips for Finding Leaks

Use the black rubber probe tip whenever possible
and hold the probe approximately 1/4 in. above the
area being tested.  Move the probe tip at about
two inches per second along seams or joints
suspected of leaking.

Start leak checking anywhere, and continue in a
logical progression through the entire system to
locate all leaks.

Set the range switch to the appropriate range to
ensure maximum sensor life.  If the leak size is
unknown, start at the large (“LG”) leak setting.
Adjust the balance control to maintain slow ticking
(three or four ticks per second) in the idle state.
(See Figure 8.)

IMPORTANT: Exposing the sensor to large

amounts of refrigerant or holding the
probe over a leak for a long period of
time will significantly shorten sensor
life.

On CFC and HCFC systems, the medium (“MED”)
range is sufficient to verify that the system is free
of leaks that require fixing.

On HFC systems, use the small (“SM”) range to
verify that the system is free of leaks that require
fixing.

Replace the filter often (as described in the

Maintenance

 section) to retain airflow and

sensitivity.

Wipe all surfaces to be tested with a clean, dry
cloth.

Check ports for moisture before inserting the
probe.

!

 CAUTION: Equipment Damage Hazard.

Submerging the probe in liquid
will damage the vacuum pump.

Summary of Contents for RLD-H10G

Page 1: ...mission Heated Diode Sensor Provides the most sensitivity available today while still detecting all halogenated refrigerant gases Halogen Selective Sensing Eliminates many sources of possible false alarms due to background contamination High Quality Air Pump Supplies constant air flow to sensor so it responds quickly to leaks also helps sensor recover quickly after exposure to refrigerant so leak ...

Page 2: ...on technology which is based on the ionization characteristics of alkali metals The alkali metals act as the cathode of a diode and the anode is heated by a DC voltage Air from the area being leak checked is pumped from the probe and into the space between the cathode and anode If halogen gas is present it creates an ionized current between the cathode and the anode which is amplified to sound a s...

Page 3: ...ch to on position See Figure 8 3 Slide range switch to small SM leak position See Figure 8 4 Check for sufficient airflow by pointing the probe tip toward the floor and observing the airflow ball in the probe See Figure 9 If the airflow is sufficient the ball will rise and float a small distance above the tip of the probe If the ball does not rise tap the probe lightly to make sure it is not stick...

Page 4: ...nd to the leak vial make sure there is liquid in the bottle If the unit still does not respond to the leak increase the sensor heat slightly by turning the sensor heat control knob on the bottom of the unit 1 16th of a turn clockwise 5 Allow a minute for the sensor to heat then repeat the calibration procedure Tips for Finding Leaks Use the black rubber probe tip whenever possible and hold the pro...

Page 5: ...emove the old airflow indicator ball 3 Insert the new airflow indicator ball into the probe 4 Reattach the probe end to the hose Figure 10 Replacing the Filter and Airflow Indicator Replacing the Sensor The sensor is a plug in unit that is thrown away when expended It needs replacement when the RLD H10G no longer responds to the leak vial even with the sensor heat control turned fully clockwise Ma...

Page 6: ...r RLD H10 602R Heat control set too low Increase the sensor heat slightly by turning the sensor heat knob on the bottom of the unit 1 16th of a turn clockwise Allow a minute for the sensor to heat then repeat calibration Audio normal probe lamp does not flash Burned out or damaged probe lamp Return unit to your Johnson Controls PENN representative Responds continuously in all leak positions Dirt i...

Page 7: ... Should it become necessary to repair your RLD H10G please contact your Johnson Controls PENN representative Warranty Repairs Units under warranty will be shipped to Johnson Controls PENN for repairs The repairs will be completed within 24 hours after receipt by the repair center and the unit will be returned to your Johnson Controls PENN representative at the same priority of shipment with which ...

Page 8: ...racy Meets SAE J1627 test requirements Probe Length Approximately 4 5 ft 1 4 m Ambient Operating Conditions 32 to 113 F 0 to 45 C 5 90 RH non condensing Ambient Storage Conditions 14 to 140 F 10 to 60 C 5 90 RH non condensing Case Vinyl Dimensions H x W x D 2 8 x 8 8 x 5 0 in 71 x 224 x 127 mm Shipping Weight 3 2 lb 1 4 kg Agency Listings UL Listed File SA9717 The performance specifications are no...

Reviews: