Trouble Shooting Guide
If something out of the normal happens during operation, please check and correct the
following points:

Symptom

Possible cause

Suggested solutions

Although the
device is on, it
does not work

1-Incorrectly connected to 

power source

2-No power
3-The AC Adapter is not connected

1-Check the plug is

properly connected

2-Check that power is on
3-Check the AC Adapter is

properly connected

The mist does
not come out

1-Not enough medication in jet

nebulizer.

2-Too much medication.
3-PVC tubing disconnected.

1-Add medication,

maximum 10 ml.

2-Remove excess

medication from cup.

3-Connect the PVC tubing.

Aerosol output
reduced (low)

1-Air filter is clogged.
2-The jet nebulizer is

dirty.

1-Replace for a clean air

filter.
2-Replace for new one.

Summary of Contents for Neb-u-Tyke RR JB0112-064

Page 1: ...future reference Caution Federal law restricts this device to sale by or on order of a physician or any other practitioner licensed by the law of the State in which he practices to use or order the use of this device MODEL JB0112 064 JB0112 064 INS LABRevB06 Neb u TykeRR COMPRESSOR NEBULIZER ...

Page 2: ... DC Class II Type B Read the instructions Contents Introduction 3 Important Safeguards 3 Features 4 Unit Assembly 5 Operating Instructions 6 Cleaning and Maintenance 7 Spare Parts Provided 7 Technical Specifications 8 Trouble Shooting Guide 9 ...

Page 3: ... unsuitable and thus dangerous The manufacturer can not be considered responsible of the damages caused by unsuitable incorrect or irrational use if the unit has been used in electrical installations not in accordance with the standing safety standards 3 The following precautions are especially important The unit is not protected from liquid penetration The nebulizer must not be touched with wet h...

Page 4: ...he mouth nozzle are for personal use and they must not be shared in order to avoid any infection or contagion 12 Children should use this unit only under adult supervision 13 This unit must be repaired only by authorized personal NOTE This unit employs a thermal limiter that will shut off the entire product should the motor overheat Jet nebulizer PVC tubing Filter Chamber ON OFF Switch Air outlet ...

Page 5: ...5 Unit Assembly IMPORTANT Do not pour more than 10 ml ...

Page 6: ...F switch to start the treatment 8 Turn the unit off and unplug it once the treatment has finished 9 Clean the jet nebulizer and the devices as indicated in Cleaning section 10 Store the jet nebulizer and the devices in the unit s storage compartment Filter change 1 Filter should be changed every 6 months or sooner if the filter turns completely gray in color 2 Remove filter cap by grasping it firm...

Page 7: ...e nebulizer parts with a towel because this could contaminate them NOTICE It is recommended to clean all accessories after each use to avoid possible infection risk due to contaminated medication Nebulize only medications prescribed by the physician and follow the physician s indications Do not use balsam or aromatic substances Use a dry cloth to clean the out part of the unit The jet nebulizer an...

Page 8: ...less than 5 microns Medication outlet 0 33 ml min or greater Liter flow range 6 8 l min Approximate maximum pressure 32 5 psi Weight 3 75 lbs Unit dimensions 6 7 x 5 5 x 4 Operation Humidity and Temperature 41º to 104º F 30 to 95 RH Storage Humidity and Temperature 14º to 104º F 10 to 95 RH Atmospheric pressure range for carriage and storage 500 1060 mb The specifications are subject to change wit...

Page 9: ...dapter is not connected 1 Check the plug is properly connected 2 Check that power is on 3 Check the AC Adapter is properly connected The mist does not come out 1 Not enough medication in jet nebulizer 2 Too much medication 3 PVC tubing disconnected 1 Add medication maximum 10 ml 2 Remove excess medication from cup 3 Connect the PVC tubing Aerosol output reduced low 1 Air filter is clogged 2 The je...

Page 10: ...overed by this limited warranty THIS LIMITED WARRANTY IS IN LIEU OFALL OTHER WARRANTIES WHETHER ORAL WRITTEN EXPRESS OR IMPLIED THERE ARE NO WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE EXCEPT FOR THOSE IMPLIED WARRANTIES ARISING BY OPERATION OF LAW WHICH ARE LIMITED IN DURATION TO THE 1 YEAR TERM OF THIS WARRANTY GRAHAM FIELD S LIABILITY WHETHER BASED ON CONTRACT TORT WARRANT...