Classification
The TLS Tunable Laser Source consists of an exposed metal chassis that is
connected directly to earth via a power cord and, therefore, is classified as a
Class 1 instrument.
Laser Specifications
The laser contained in the TLS source is Class 1 laser as specified under the laser
classification of the US Food and Drug Administration (FDA) Center for Devices
and Radiological Health (CDRH). Laser specifications are provided in Table 2
below.
Table 2: Specifications
Parameter
Single-mode (SM)
Wavelength
1260 to 1630 nm
Class
1
Fiber Type
9/125 µm SM
Maximum Accessible
Emission Level
-15 dBm
Maximum Output
Power
-1 dBm
Numerical Aperture
0.13
Effective Numerical
Aperture
0.20
Important Safety Information
Laser Hazards
Warning
Never look into the end of an optical cable connected to an
optical output device that is operating. Laser radiation is
invisible, and direct exposure can severely injure the human
eye.