11 

 

Do not bend the supply cable or any other equipment with force and do not place any heavy 
or sharp objects on them (danger of fire or electric shock). 

Do not plug more than one device into one power source (danger of fire).  

Check the parameters of power source before plugging the device in.  

If other than recommended power source is used, electric circuits can be damaged or there can 
be a danger of a fire.  

It is necessary to monitor the effect of plasma discharge on skin cells during the treatment. If 
any unexpected effect appears, the tip must be taken off the skin immediately and treatment 
must be stopped immediately.  

Do not place the device at the place, where it could be unplugged from the power source by 
chance.  

 

Error Messages on the Screen: 

 

        TAPE 

 Grounding of patient

  

The  device  is  equipped  with  a  reliable  safety  system  SCS  (Safety  Control 
System), which is able to detect whether the patient is grounded.  

Warning: 

Patients must have the grounding electrode stick to their body during 

the whole time of the treatment. If the patient is nt grounded, the display shows 

the  warning  „TAPE“  and  the  generator  the  Plasma  is  disconnected 

automatically.  
 

        TErr -   

Two buttons pushed simultaneously. Error of the Operation, the message 

“TErr” 

appears.  

 

        SrvC -   Hardware error 

If a hardware error appears, the device switches off automatically and an error 

message  “SrvC”

  appears.  The  device  cannot  be  switched  on  anymore  and  it 

must be sent into an authorized service.  

 

 

Written consent, training of the staff: 

Only trained doctors can use the device.  

DANGER: 

The trained staff must make sure that the patient does not suffer from any 

of  the  contraindications  before  the  treatment  and  he  must  make  the  patient  sign  the 
informed consent. Training of the operators is provided either by the manufacturer or 
authorized  distributor.  The  trained  staff  receives  a  certificate  entitling  is  holder  to 
work  with  Jett  Plasma  Lift  Medical.  The  trained  staff  must  make  sure  that  in  his 

workplace untrained people can´t work with the device. 

 

The  trained  staff  must  continuously  follow  the  working  protocols  set  by  the 
manufacturer and must avoid following safety dangers:  

Summary of Contents for Plasma Lift Medical

Page 1: ...JPL Medical Date 9 9 2015 Jett Plasma Lift Medical User Manual Before you start using Jet Plasma Lift read this manual and make sure you understand all instructions and warnings and follow them close...

Page 2: ...the Manual 10 6 Warnings and precautions 10 7 Operation of the device 13 8 Definition of plasma 17 8 1 The nature of discharge generating 18 9 Effect on cells 18 10 Use 19 11 Types of treatment 19 12...

Page 3: ...vice and conductively connected to the device skin of a patient and maintaining the distance of 2mm between the tip and skin this constant distance is provided by a specific applicator a spark dischar...

Page 4: ...lectrodes solid gel 51 33 mm v 300612979 Flat applicator Golden applicator with pointed tip Cone applicator AC adapter Type DA12 050EU M Input 100 240V 50 60Hz Output 5VDC max 2 0A Power max 12 VA Man...

Page 5: ...e state The aim of the use are corrections removal of skin structures by the use of sparkle discharges generated by DC current Some of the aforementioned effects indications are not of a clinical char...

Page 6: ...this is that the energy absorbed by body can be in cases of some applications used for targeted destruction of skin cells At the highest levels 6 7 and 8 this device is used for electro fulguration e...

Page 7: ...used in classical surgery dermatology gynecology or dental medicine At medium intensities 5 6 when there is no considerable destruction of skin cells the device is used for the treatment of small vein...

Page 8: ...device it must not be disposed of in communal waste Graphic sign Double insulation Graphic sign Use indoors Graphic sign Protection from a lightning Graphic sign ground galvanic connection with the de...

Page 9: ...r storage and transport Graphic sign Range of pressure min 50 kPa max 106 kPa Graphic sign Humidity limitations min 30 max 75 Graphic sign Protect from rain Labels at the rear side of the device Flat...

Page 10: ...explanations before you start to operate the device PRECAUTION introductory precautions All parts of the device must be connected in their places in the designed positions Use of a power adaptor other...

Page 11: ...stick to their body during the whole time of the treatment If the patient is nt grounded the display shows the warning TAPE and the generator the Plasma is disconnected automatically TErr Two buttons...

Page 12: ...rounding Caution Before the treatment begins the patient must be grounded by a grounding disposable electrode which is stuck on the right shoulder blade Do not touch the tip Caution Neither operator n...

Page 13: ...the device can destroy skin cells at the intensity of 6 7 and 8 Before every application make sure twice that you have set the right intensity That means that if you do the treatment of couperosis or...

Page 14: ...ther side to the disposable grounding electrode The message TAPE disappears from the display and the lowest intensity level 1 appears c Setting the spark discharge intensity After switching the device...

Page 15: ...ic applicator has the same use but it does not touch skin it touches only the indifferent gel 2 Golden applicator with precise tip It is designed for face treatments and precise removals of skin pertu...

Page 16: ...ce i Cosmetic treatment Lower intensities 1 5 Approach the tip of the device 1 to a treated place protected by an even layer of conductive indifferent gel or by a layer of a conductive serum for examp...

Page 17: ...ectrons created by the separation of electrons from the electron shell of gases or by ionization Plasma can be also defined as the fourth form of matter In this case plasma is quasineutral and low tem...

Page 18: ...ong the membrane and the electric voltage of the membrane is changed The Membrane potential is created and influenced by potassium and sodium cations It is difficult for sodium cations to pass through...

Page 19: ...ted The spark channel generated by DC voltage is very slim approx 1 mm2 and therefore the treatment and the targeted destruction of cells is very precise The treatment of skin by spark discharge is a...

Page 20: ...1st and the 2nd treatment there should be 3 free days after every other treatment a longer free period of 1 week is recommended One treatment including preparation time and aftercare takes 1 hour Afte...

Page 21: ...ow treatment Move the tip of the device firmly slowly and carefully in the area of wrinkles 1 2 mm above the gel as you have been taught at the training If bigger areas are treated use the flat applic...

Page 22: ...e at the tip to zero after switching off after 1 s Size 245 mm 45 mm Weight approx 350 g Range of relative air humidity 30 75 Range of environment temperature 10 to 40 o C Range of atmospheric pressur...

Page 23: ...nd humidity Store unplugged from power supply Regular maintenance cleaning a periodical safety technical checks in given time Maintenance the operators are obliged to follow all the instructions in th...

Page 24: ...of electrical or electronic devices may contain substance that may be harmful for the environment or health Jett Plasma Lift Medical is included in the electric waste and after the end of its lifespa...

Page 25: ...420 602 766 759 00420 725 385 074 e mail nespor compexbrno cz export compexbrno cz The manufacturer has a system of authorized service centers For more information see www jettcosmetics cz 13 5 The De...

Page 26: ...or A 0001 15 DATE OF MANUFACTURE 15 2 2015 Compex spol s r o Palack ho t da 105 612 00 Brno Czech Republic tel 00420 602 766 759 www jettmedical cz Golden applicator with a precise tip B 0001 15 DATE...

Page 27: ...27 C 0001 15 DATE OF MANUFACTURE 15 2 2015...

Reviews: