Installation, Operation, & Maintenance Instructions
IOM J500.35
2021.04
JMT
● +1.440.572.1500
4
Section 1: Unpacking
1.1 Storage and Handling
1. Upon receipt of materials, open crates / boxes and inspect all materials to ensure no damage
has occurred during transit. If damage is noted, file claim with Clark-Reliance or the applicable
freight carrier, depending on shipping terms.
2. Replace and fasten down lid of crate / box.
3. Materials shall remain inside of the crate / box until it is ready to be installed.
4. Store all materials in a dry, clean, indoor environment, protected from the elements, flooding,
corrosive fumes or other physical damage.
5. Materials should be stored in a place where they can neither fall nor be struck by falling
objects.
6. Storage area should be climate (humidity) controlled environment to prevent condensation
and corrosion.
a. Storage temperature for shipments that include transmitters should be 40°F to 104°F
(4.5°C to 40°C).
7. Outdoor storage is not recommended. Damage caused by outdoor storage or other negligent
storage is not covered by Clark-Reliance Warranty.
a. JMT Transmitters ship with NPT type plugs to prevent water ingress. These plugs
should not be considered watertight, nor are meant to prevent ingress from exposure
to weather.
1.2 Factory Mounted on Jerguson
®
Magnicator
®
**Do not discard any packaging until unit is fully installed and operational**
Refer to Jerguson Magnicator IOM M500.01 for more information
If your JMT Magnetostrictive comes pre-installed on a Jerguson Magnicator, please perform the
following checks to ensure there was no damage during shipping.
1. Ensure brackets are secure and transmitter head and sensor probe are free from damage,
including bends of impact damage.
2.
Check the ‘4mA’ and ‘20mA’ stickers on the sensor probe are in place and line-up with the lower
and upper side process connections. If your Magnicator is a configuration that does not have
either one or both side connections or has a measuring range that does not match either one or
both side process connections, ensure the
‘4mA’ lines up with the lowest point of the measuring
range
(whether inches, %, mm or otherwise) on the flag indicator and the ‘20mA’ corresponds
with the span, i.e., full measuring range.