JEOL JEM-2100 HR Standard Operating Procedure Download Page 7

20 November 2010, R. Ubic                                            

7

 

Filament Ramping

 

 

Make sure the 

Normal Setup

 button is selected in the Filament Control Speed. 

 

The filament should be set to ramp to some percentage (typically 60% - 70%) less than 

the maximum limit.  Do not alter this number or the filament maximum limit. 

 

Press the filament ON button in the High Voltage Control window (or the 

Beam

 button 

on the top left of the left-hand control panel). 

 

The Filament window appears.  While it is open you cannot access any other windows. 

 

Alignment Procedure

 

 

Finding the beam

 

 

The last user should have left the microscope at a magnification of about 50k and with 

all the apertures out.  If you can't find the beam, the most likely reason is that your 

sample is in the way.  Try moving it around.  DO NOT move apertures or use beam shift 

knobs to find the beam! 

 

If you can't find the beam, try: 

 

1. Check that the beam current has reached its normal value (½ kV ± 10%). 

2. Check that all the apertures are out. 

3. Check that the spot size is set to 1 (use 

SPOT SIZE

 knob on left-hand panel). 

 

4. Decrease the magnification and spread the beam. 

5. Go to 

LOW MAG

 mode (button on right-hand panel) and spread the beam with 

the 

BRIGHTNESS

 knob (on left-hand panel). 

6. Move your sample around with the trackball. 

7. If you still can't find it, turn off the filament and remove the sample holder to the 

park position.  Turn on the filament again and repeat the above steps to find the 

beam.  If you find it, center it, then shut off the filament, re-insert your sample, 

and begin again. 

8. Try loading a previous alignment (Maintenance - Alignment - Load Alignment 

File) 

9. If you still can't find the beam, get assistance from Rick Ubic. 

 

Gun Alignment

 

 

Find a hole in the sample. 

 
For initial alignment, set the 

α

-selector to 3.  In general, for magnifications less than 

100k, 

α

 = 3 is appropriate.  Setting 2 is recommended for magnifications above 100k, 

and setting 1 is recommended for magnifications of 400k and above.  

 

 

Summary of Contents for JEM-2100 HR

Page 1: ...0 HR TRANSMISSION ELECTRON MICROSCOPE TEM EDS Chemical analysis Post column annular dark field detector Tomography unit for 3D images Cryogenic temperature control down to 77 K Point Resolution 2 3 Li...

Page 2: ...ow Cooling Water Suspension The high voltage power and lens power will shut down and the vacuum system will go into the protected state Upon resumption of cooling water the microscope must be restarte...

Page 3: ...mn on left hand side Check to make sure that the gun gas pressure is 0 3 MPa bottom round dial gauge behind grill at base of column on left hand side Check to make sure that the column pressure gauge...

Page 4: ...ewar with liquid nitrogen For operation at 100 200 kV Use the and buttons to increase the kV setting to 100 kV The Step option allows you to control the step size of the changes Press the HT ON button...

Page 5: ...tion and secure it with the retainer screws Open the clamp on the holder again with the cartridge handling tool and re insert the cartridge carefully locating it on the pin Double Tilt Beryllium Holde...

Page 6: ...for the green light to appear DO NOT ROTATE THE HOLDER YET If using the double tilt holder place the y tilt control cable on top of the goniometer for support Do not let the cable dangle Turn the hol...

Page 7: ...e beam current has reached its normal value kV 10 2 Check that all the apertures are out 3 Check that the spot size is set to 1 use SPOT SIZE knob on left hand panel 4 Decrease the magnification and s...

Page 8: ...maximize the current density brightness on the screen Press F4 again to de select the gun assignments Condenser Astigmatism Correction Press the COND STIG button left hand panel and use the DEF STIG k...

Page 9: ...T to shut this wobbler off Condenser Lens Deflection Coil Adjustment Tilt Purity Focus the beam Click on Tilt in the Compensator portion of the Alignment Panel for Maintenance window Click on TiltX in...

Page 10: ...er portion of the Alignment Panel for Maintenance window Unify the resulting two spots using the DEF STIG X knob If the spot moves off screen reduce the camera length and continue adjusting If the spo...

Page 11: ...AGE WOBB Y button right hand panel Unify the oscillating image with the OBJ FOCUS knobs right hand panel If the image cannot be unified try correcting for objective astigmatism above Press the IMAGE W...

Page 12: ...NESS knob Adjust the DIFF FOCUS knob right hand panel to obtain the sharpest pattern Center the pattern by pressing PLA on the left hand console and adjusting position with ShiftX and ShiftY knobs To...

Page 13: ...indow 4 Move the cross cursor to the center of the screen by clicking on the Center button in the Image Control window 5 Turn the BRIGHTNESS knob fully clockwise 6 Obtain a caustic spot using the OBJ...

Page 14: ...justing the DEF STIG knobs If the Ronchigram is not circular or if adjusting the OBJ FOCUS knobs make it elliptical press the COND STIG button left hand panel and adjust the DEF STIG knobs accordingly...

Page 15: ...is perfectly fine or if you want to get more detail in the tomogram you can pick 1 degrees Pay attention to the mag and the at start and when needed track at magnification Rule of thumb here is the la...

Page 16: ...een if this window is not still open Wait until filament is completely off before extracting the holder Do NOT remove the holder while the filament is turning off Remove sample from holder DO NOT STOR...

Reviews: