background image

20 November 2010, R. Ubic                                            

4

 

Raising the Voltage

 

All references to the screen refer to the left-hand monitor unless otherwise stated. 

 

If the “Controller for JEM-2100/HR” software is not currently running, double click on the 

TEMCON icon on the desktop of the left-hand monitor. 

 

The beam current should be zero and the high voltage (HT) switched off unless the 

microscope has recently been used, in which case the HT will be left on and beam 

current non-zero. 

 

If the anti-contamination device heater is in place, carefully unplug it and remove it. 

 

Click on the HT button at the top of the screen. 

 

While you are waiting for the high-voltage to stabilize, fill the anticontamination device 
(ACD) dewar with liquid nitrogen. 

 

For operation at 100 - 200 kV: 

 

Use the 

 and 

 buttons to increase the kV setting to 100 kV.  The 

Step

 option allows 

you to control the step size of the changes. 

 

Press the HT 

ON

 button and wait for the voltage to stabilize at 100 kV. 

 

If the desired voltage is greater than 100 kV:  

 

Click on the 

More

 box to obtain more options. 

 

In the Auto HT section, set the 

Target

 voltage to the desired value.  Again, the 

step controller will allow you to change the size of voltage steps. 

 

Set the 

Step

 size to 0.1 and the 

Time/Step

 to 3. 

 

Press the 

Start

 button in the Auto HT section.  If the target voltage is 200 kV, 

then this will take about 50 minutes. 

 

Once the HT stabilizes, the 

beam current

 (top left of screen) should be about 

half the voltage (in kV) 

µ

A (or about 101 

µ

A at 200 kV).  

If it is > 105 

µ

A at       

200 kV, contact Rick Ubic or Steve Letourneau and do not attempt to use the 
microscope.

 

 

Summary of Contents for JEM-2100 HR

Page 1: ...0 HR TRANSMISSION ELECTRON MICROSCOPE TEM EDS Chemical analysis Post column annular dark field detector Tomography unit for 3D images Cryogenic temperature control down to 77 K Point Resolution 2 3 Li...

Page 2: ...ow Cooling Water Suspension The high voltage power and lens power will shut down and the vacuum system will go into the protected state Upon resumption of cooling water the microscope must be restarte...

Page 3: ...mn on left hand side Check to make sure that the gun gas pressure is 0 3 MPa bottom round dial gauge behind grill at base of column on left hand side Check to make sure that the column pressure gauge...

Page 4: ...ewar with liquid nitrogen For operation at 100 200 kV Use the and buttons to increase the kV setting to 100 kV The Step option allows you to control the step size of the changes Press the HT ON button...

Page 5: ...tion and secure it with the retainer screws Open the clamp on the holder again with the cartridge handling tool and re insert the cartridge carefully locating it on the pin Double Tilt Beryllium Holde...

Page 6: ...for the green light to appear DO NOT ROTATE THE HOLDER YET If using the double tilt holder place the y tilt control cable on top of the goniometer for support Do not let the cable dangle Turn the hol...

Page 7: ...e beam current has reached its normal value kV 10 2 Check that all the apertures are out 3 Check that the spot size is set to 1 use SPOT SIZE knob on left hand panel 4 Decrease the magnification and s...

Page 8: ...maximize the current density brightness on the screen Press F4 again to de select the gun assignments Condenser Astigmatism Correction Press the COND STIG button left hand panel and use the DEF STIG k...

Page 9: ...T to shut this wobbler off Condenser Lens Deflection Coil Adjustment Tilt Purity Focus the beam Click on Tilt in the Compensator portion of the Alignment Panel for Maintenance window Click on TiltX in...

Page 10: ...er portion of the Alignment Panel for Maintenance window Unify the resulting two spots using the DEF STIG X knob If the spot moves off screen reduce the camera length and continue adjusting If the spo...

Page 11: ...AGE WOBB Y button right hand panel Unify the oscillating image with the OBJ FOCUS knobs right hand panel If the image cannot be unified try correcting for objective astigmatism above Press the IMAGE W...

Page 12: ...NESS knob Adjust the DIFF FOCUS knob right hand panel to obtain the sharpest pattern Center the pattern by pressing PLA on the left hand console and adjusting position with ShiftX and ShiftY knobs To...

Page 13: ...indow 4 Move the cross cursor to the center of the screen by clicking on the Center button in the Image Control window 5 Turn the BRIGHTNESS knob fully clockwise 6 Obtain a caustic spot using the OBJ...

Page 14: ...justing the DEF STIG knobs If the Ronchigram is not circular or if adjusting the OBJ FOCUS knobs make it elliptical press the COND STIG button left hand panel and adjust the DEF STIG knobs accordingly...

Page 15: ...is perfectly fine or if you want to get more detail in the tomogram you can pick 1 degrees Pay attention to the mag and the at start and when needed track at magnification Rule of thumb here is the la...

Page 16: ...een if this window is not still open Wait until filament is completely off before extracting the holder Do NOT remove the holder while the filament is turning off Remove sample from holder DO NOT STOR...

Reviews: