background image

20 November 2010, R. Ubic                                            

15

 

Tomography 

 

Insert the holder and bring up the beam at 3000x. Make sure the beam is spread out but 

don't go all the way to fully overfocus with CL3. 

 

Acquire a trial image with SerialEM. If what you see looks contrast-rich and has many 

features, start 'eucentric - rough' from the task menu. 

 

Once this completes, locate your ROI and set the magnification and brightness (you can 

move the stage and set the brightness all through SerialEM). 

 

Once you have an ROI for tomography run 'eucentric - fine' from the task menu. 

 

Walk up the stage to 53 degrees using 'walk up' from the task menu. This starting angle 

may be higher if you use a High Tilt Retainer and no in-gap OL aperture in the beam 

path. 

 

From the 'tilt series' menu choose 'Setup/start' to evaluate your tomography parameters. 

Make sure that the angles are properly specified and pick your basic increment (1.5 or 2 

is perfectly fine, or if you want to get more detail in the tomogram you can pick 1 

degrees). Pay attention to the mag and the 'at start and when needed track at 

magnification'. Rule of thumb here is the latter should probably be ~1/3rd of the imaging 

magnification. 

 

Keep the beam intensity for the moment constant. 

 

Most biological samples do fine with a focus of -0.35 micron, so do most materials 

samples. For cryo you need more, say 3-6 microns. If the ROI has features that only 

show with a lot of defocus, you can add this as an offset to the autofocus (remember 

that underfocus is negative!). 

 

For initial action you probably should have a trial image in focus before you bring up this 

window, so 'align to image now in Buffer A' should be on. 

 

Leave tracking unchanged for the moment. 

 

Hit 'GO' and sit back. Remember there are 'End' and 'STOP' on the camera and macro 

controls panel that will halt the tilt series acquisition. 

 

Once the series is done, terminate it from the 'Tilt Series' menu and close the file and do 

not leave 90° rotated copies in various buffers. 

 

You can run up the tomogram in etomo. 

 

 

 

Summary of Contents for JEM-2100 HR

Page 1: ...0 HR TRANSMISSION ELECTRON MICROSCOPE TEM EDS Chemical analysis Post column annular dark field detector Tomography unit for 3D images Cryogenic temperature control down to 77 K Point Resolution 2 3 Li...

Page 2: ...ow Cooling Water Suspension The high voltage power and lens power will shut down and the vacuum system will go into the protected state Upon resumption of cooling water the microscope must be restarte...

Page 3: ...mn on left hand side Check to make sure that the gun gas pressure is 0 3 MPa bottom round dial gauge behind grill at base of column on left hand side Check to make sure that the column pressure gauge...

Page 4: ...ewar with liquid nitrogen For operation at 100 200 kV Use the and buttons to increase the kV setting to 100 kV The Step option allows you to control the step size of the changes Press the HT ON button...

Page 5: ...tion and secure it with the retainer screws Open the clamp on the holder again with the cartridge handling tool and re insert the cartridge carefully locating it on the pin Double Tilt Beryllium Holde...

Page 6: ...for the green light to appear DO NOT ROTATE THE HOLDER YET If using the double tilt holder place the y tilt control cable on top of the goniometer for support Do not let the cable dangle Turn the hol...

Page 7: ...e beam current has reached its normal value kV 10 2 Check that all the apertures are out 3 Check that the spot size is set to 1 use SPOT SIZE knob on left hand panel 4 Decrease the magnification and s...

Page 8: ...maximize the current density brightness on the screen Press F4 again to de select the gun assignments Condenser Astigmatism Correction Press the COND STIG button left hand panel and use the DEF STIG k...

Page 9: ...T to shut this wobbler off Condenser Lens Deflection Coil Adjustment Tilt Purity Focus the beam Click on Tilt in the Compensator portion of the Alignment Panel for Maintenance window Click on TiltX in...

Page 10: ...er portion of the Alignment Panel for Maintenance window Unify the resulting two spots using the DEF STIG X knob If the spot moves off screen reduce the camera length and continue adjusting If the spo...

Page 11: ...AGE WOBB Y button right hand panel Unify the oscillating image with the OBJ FOCUS knobs right hand panel If the image cannot be unified try correcting for objective astigmatism above Press the IMAGE W...

Page 12: ...NESS knob Adjust the DIFF FOCUS knob right hand panel to obtain the sharpest pattern Center the pattern by pressing PLA on the left hand console and adjusting position with ShiftX and ShiftY knobs To...

Page 13: ...indow 4 Move the cross cursor to the center of the screen by clicking on the Center button in the Image Control window 5 Turn the BRIGHTNESS knob fully clockwise 6 Obtain a caustic spot using the OBJ...

Page 14: ...justing the DEF STIG knobs If the Ronchigram is not circular or if adjusting the OBJ FOCUS knobs make it elliptical press the COND STIG button left hand panel and adjust the DEF STIG knobs accordingly...

Page 15: ...is perfectly fine or if you want to get more detail in the tomogram you can pick 1 degrees Pay attention to the mag and the at start and when needed track at magnification Rule of thumb here is the la...

Page 16: ...een if this window is not still open Wait until filament is completely off before extracting the holder Do NOT remove the holder while the filament is turning off Remove sample from holder DO NOT STOR...

Reviews: