7.7
Unable to set blank
Possible cause
Solution
Blank contaminated.
Re-make the blank and re-check.
Burner encrusted with salt.
Clean the burner as defined in Section 6.4, Six
monthly maintenance.
Ensure that the cleaning procedure is carried out
more frequently.
7.8
No reading on display
Possible cause
Solution
Incorrect filter selected.
Select correct filter.
Calibration standard incorrectly
made.
Re-make standard.
Nebuliser completely blocked.
Check the nebuliser performance as defined in
Section 6.2, Weekly maintenance.
Section 8
Spares
8.1
Minor Spares Kit (PFP7/MINKIT)
Description
Part code
Quantity
Nebuliser
500019
1
Cleaning Wire (3x150mm)
I/500194
1
Cleaning Solution
025016
2
Ignition Probe Assembly
PFP7/IPA
1
Mk2 Small Knob (Blank & Fine) Assembly
PFP7/SKB
1
Mk2 Large Knob (d.p., Coarse & Filter Select)
Assembly
PFP7/LKB
1
Mk2 Fuel Knob Assembly
PFP7/FKB
1
Fuse Kit
PFP7/FKIT
1
Drain Trap Kit
500551
1
Ignition Switch
017025
1
Power Switch
017050
1
Nebuliser to Solution Tube
I/500193
500mm
Air Regulator to Nebuliser Tube
023009
500mm
Air Regulator to Nebuliser Tube (Plastic Olive)
026023
1
Air Regulator to Nebuliser Tube (Brass Nut)
026007
1
Air Regulator to Nebuliser Tube (Coupling to Air
Regulator)
026024
1
Allen Key
060196
1
Foot Assembly
PFP7/FA
1
500 795/Rev E/03-17
Page 4