27
IX. SPECIFICATIONS
Conductivity / TDS
RANGE
Basic Cell K
Resolution
Accuracy
0 to 9.999 uS/ppm
0.01
0.001uS/ppm
0.5%
±
1 LSD
1 to 99.99 uS/ppm
0.01
0.01uS/ppm
0.5%
±
1 LSD
10 to 300.0 uS/ppm
0.01
0.1uS/ppm
0.5%
±
1 LSD
0.0 to 99.99 uS/ppm
0.10
0.01uS/ppm
0.5%
±
1 LSD
10 to 999.9 uS/ppm
0.10
0.1uS/ppm
0.5%
±
1 LSD
0.1 to 3.000 mS/ppt
0.10
0.001mS/ppt
0.5%
±
1 LSD
0 to 999.9 uS/ppm
1.00
0.1 uS/ppm
0.5%
±
1 LSD
0.1 to 9.999 mS/ppt
1.00
0.001mS/ppt
0.5%
±
1 LSD
1.0 to 30.00 mS/ppt
1.00
0.01mS/ppt
0.5%
±
1 LSD
0 to 9.999 mS/ppt
10.0
0.001mS/ppt
0.5%
±
1 LSD
1 to 99.99 mS/ppt
10.0
0.01mS/ppt
0.5%
±
1 LSD
10 to 300.0 mS/ppt
10.0
0.1mS/ppt
0.5%
±
1 LSD
Temperature
Range Resolution
Accuracy
-10.0 to 120.0 °C 0.1 °C
±0.3 °C
Conductivity / TDS
TDS Factor Range
0.300
to
0.999
(user
selectable)
Reference Temperature
10°C to 29°C (user selectable)
Temperature Coefficient
0.00
to
4.99%
(user
selectable)
Temperature Compensation
Automatic
Temperature
Temperature sensor
Thermistor, 10 k
at 25°C