2

GENERAL INTRODUCTION   

 

 

Thank you for selecting the 6010N meter. The 6010N is a precision 
tool that measure pH, ORP and temperature. A built-in 
microprocessor stores, calculates and compensates for all 
parameters related to pH determinations including pH electrode 
temperature characteristics, electrode slope deviations and buffer 
solutions. 
 
This unit has a waterproof IP65 case. The touch mode keys are 
highly reliable with tactile and audio feedback. This meter can 
operated with one 9V battery, typical battery life is 1000 hours. 
Re-calibration is not required when power is turned on again. 
 
The front of the meter has a large LCD that displays pH or ORP and 
temperature simultaneously along with user prompts and mode 
indicators. The unit prompts the user through calibration and 
measurement procedures. 
 
An AUTOLOCK feature for both pH and ORP measurements 
enables the unit to automatically sense the end point and “ HOLD ” 
the display to indicate the end point value of a measurement. 
AUTOLOCK and user prompts help eliminate most errors in 
determining pH and mV values, resulting in precise, repeatable and 
error-free measurements. The 6010N can also be used in 
non-AUTOLOCK mode.   
 
The model 6010N is available with pH, ORP and ATC (Automatic 
Temperature Compensation) probes. Other features include 
electrode offset recognition, electrode slope recognition, electrode 
efficiency display, built-in buffer coefficients, automatic or manual 
temperature compensation, long battery life, and 50/60Hz AC noise 
rejection. This meter is user-friendly, for field, industrial and 
laboratory applications. 
 

INITIAL INSPECTION     

 

 

Carefully unpack the unit and accessories. Inspect for damages 
made in shipment. If any damage is found, notify your 

Jenco

 

representative immediately. All packing materials should be saved 
until satisfactory operation is confirmed. 
 

WATER PROOF   

 

 

Though the 6010N meter is housed in a watertight case, 

DO NOT

 

use it underwater. The watertight case prevents permanent damage 
to the unit if accidentally dropped into non-corrosive solutions.   

Summary of Contents for 6010N

Page 1: ...Operation Manual Hand held pH ORP Temperature Meter 6010N...

Page 2: ...OLOCK mode 8 d Calibration with manual temperature compensation in the pH NON AUTOLOCK mode 9 C pH Measurements 10 a Measurement with an ATC Temp probe in the pH AUTOLOCK mode 10 b Measurement with ma...

Page 3: ...HOLD the display to indicate the end point value of a measurement AUTOLOCK and user prompts help eliminate most errors in determining pH and mV values resulting in precise repeatable and error free me...

Page 4: ...NSTALLING THE BATTERIES The 6010N meter is packaged with one 9V battery required for operation To insert the batteries into the meter follow the procedure outlined below Figure 1 Battery compartment 1...

Page 5: ...he electrode 3 HOLD This will indicate that the reading is frozen during AUTOLOCK mode 8 pH mV Unit and mode indicators 4 ATC MAN ATC indicator will be displayed if a temperature probe is connected ot...

Page 6: ...it Pressing and holding the STAND key while turning on the power will change the buffer set to the other available buffer set MEA EFF The key is used to bring the unit out of the AUTOLOCK condition wh...

Page 7: ...and temperature readings with the pH icon on but AUTOLOCK icon off 4 mV mode The unit will display mV and temperature readings with the mV icon on but AUTOLOCK icon off OPERATIONAL PROCEDURES A Buffe...

Page 8: ...temperature reading to stabilize then press and hold STAND key for 2 seconds to calibrate The WAIT icon will flash until the unit detects a stable reading Once the unit calibrates the first point the...

Page 9: ...he pH probe in distilled water and immerse it in the second buffer solution either 4 00 4 01 or 9 18 10 01 then press SLOPE key to calibrate The WAIT icon will flash until the unit detects a stable re...

Page 10: ...en the efficiency value is over than 75 d Calibration with manual temperature compensation in the pH NON AUTOLOCK mode 1 Turn the unit on Press CLEAR key for 5 seconds all LCD elements will lit up The...

Page 11: ...and ATC temp probe with distilled water and immerse them in the sample to be measured Remove any air bubbles trapped around the probe by shaking or stirring the probe 4 Press the MEA EFF key The WAIT...

Page 12: ...the pH NON AUTOLOCK mode for measurements c Measurement with an ATC Temp probe in the pH NON AUTOLOCK mode 1 Connect the pH electrode to the BNC connector and the ATC Temp probe to the ATC Temp connec...

Page 13: ...Press the MEA EFF key The WAIT icon will start flashing The unit is waiting for a stable reading The meter will track the mV value as sensed by the ORP electrode 5 When the WAIT icon disappears and th...

Page 14: ...0 4 01 6 98 9 46 4 01 7 11 10 32 5 4 00 6 95 9 39 4 01 7 08 10 25 10 4 00 6 92 9 33 4 00 7 06 10 18 15 4 00 6 90 9 28 4 00 7 03 10 12 20 4 00 6 88 9 23 4 00 7 01 10 06 25 4 00 6 86 9 18 4 01 7 00 10...

Page 15: ...rrect 3 pH electrode SLOPE is off by more than 30 of ideal SLOPE 1 Allow sufficient time for the electrode and ATC Temp probe to stabilize re press SLOPE key to continue the calibration procedure 2 Ch...

Page 16: ...ecognition 90 mV at pH 7 00 or 6 86 pH Electrode SLOPE recognition 30 at pH 4 00 4 01 9 18 and 10 01 Input impedance 1012 Temperature sensor Thermistor 10 k at 25 C Power 9Volt battery Battery Life 10...

Page 17: ...product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis RETURN OF ITEMS Authorization must be obtained from one of our representatives befor...

Page 18: ...coi com Website www jencoi com Jenco Electronics Inc 1F NO 11 Lane 370 Sec 6 Zhongxiao E Rd Nangang Dist Taipei Taiwan TEL 886 2 2782 3226 FAX 886 2 2782 3234 Shanghai Jenco Instruments Ltd 18 Wang Do...

Reviews: