JENCO 6010N Operation Manual Download Page 13

 

12

4. 

Allow sufficient time for the display to stabilize. The instrument 
will display the pH value of the sample at the set sample 
temperature. 

 

D.  Temperature Measure 

The 6010N can measure temperature independently with the 
ATC/temp probe without using the pH electrode. Place the 
ATC/temp probe in the sample. The unit will display the measured 
temperature. 

 

E.  mV Measurements 

a. 

Measurement in the mV-AUTOLOCK mode. 

 
1.  Connect the combination ORP electrode to the BNC connector 

of the unit. 

 
2.  Press “MODE” key until “mV” icon and “AUTO” icon lit up.   
 
3.  Rinse electrode with distilled water and immerse it in sample to 

be measured.   

 
4.  Press the “MEA./EFF.” key. The “WAIT” icon will start flashing. 

The unit is waiting for a stable reading. The meter will track the 
mV value as sensed by the ORP electrode. 

 
5.  When the “WAIT” icon disappears and the “HOLD” icon is 

displayed, the reading is then locked and will not respond to 
further changes from the sample. The mV value is the sample 
reading. 

 

[

Note: 

For samples that are inherently unstable, the unit will 

not AUTOLOCK. In this case, use the mV NON- AUTOLOCK 
mode for measurements.] 

 

b. 

Measurement in the mV NON-AUTOLOCK mode. 

 
1. 

Connect the combination ORP electrode to the BNC connector 
of the unit. 

 
2. 

Press “MODE” key until “mV” icon lit up.   

 
3. 

Rinse electrode with distilled water and immerse it in sample to 
be measured.   

 
4. 

Allow sufficient time for the display to stabilize. The instrument 
will display the mV value of the sample. 

 

Summary of Contents for 6010N

Page 1: ...Operation Manual Hand held pH ORP Temperature Meter 6010N...

Page 2: ...OLOCK mode 8 d Calibration with manual temperature compensation in the pH NON AUTOLOCK mode 9 C pH Measurements 10 a Measurement with an ATC Temp probe in the pH AUTOLOCK mode 10 b Measurement with ma...

Page 3: ...HOLD the display to indicate the end point value of a measurement AUTOLOCK and user prompts help eliminate most errors in determining pH and mV values resulting in precise repeatable and error free me...

Page 4: ...NSTALLING THE BATTERIES The 6010N meter is packaged with one 9V battery required for operation To insert the batteries into the meter follow the procedure outlined below Figure 1 Battery compartment 1...

Page 5: ...he electrode 3 HOLD This will indicate that the reading is frozen during AUTOLOCK mode 8 pH mV Unit and mode indicators 4 ATC MAN ATC indicator will be displayed if a temperature probe is connected ot...

Page 6: ...it Pressing and holding the STAND key while turning on the power will change the buffer set to the other available buffer set MEA EFF The key is used to bring the unit out of the AUTOLOCK condition wh...

Page 7: ...and temperature readings with the pH icon on but AUTOLOCK icon off 4 mV mode The unit will display mV and temperature readings with the mV icon on but AUTOLOCK icon off OPERATIONAL PROCEDURES A Buffe...

Page 8: ...temperature reading to stabilize then press and hold STAND key for 2 seconds to calibrate The WAIT icon will flash until the unit detects a stable reading Once the unit calibrates the first point the...

Page 9: ...he pH probe in distilled water and immerse it in the second buffer solution either 4 00 4 01 or 9 18 10 01 then press SLOPE key to calibrate The WAIT icon will flash until the unit detects a stable re...

Page 10: ...en the efficiency value is over than 75 d Calibration with manual temperature compensation in the pH NON AUTOLOCK mode 1 Turn the unit on Press CLEAR key for 5 seconds all LCD elements will lit up The...

Page 11: ...and ATC temp probe with distilled water and immerse them in the sample to be measured Remove any air bubbles trapped around the probe by shaking or stirring the probe 4 Press the MEA EFF key The WAIT...

Page 12: ...the pH NON AUTOLOCK mode for measurements c Measurement with an ATC Temp probe in the pH NON AUTOLOCK mode 1 Connect the pH electrode to the BNC connector and the ATC Temp probe to the ATC Temp connec...

Page 13: ...Press the MEA EFF key The WAIT icon will start flashing The unit is waiting for a stable reading The meter will track the mV value as sensed by the ORP electrode 5 When the WAIT icon disappears and th...

Page 14: ...0 4 01 6 98 9 46 4 01 7 11 10 32 5 4 00 6 95 9 39 4 01 7 08 10 25 10 4 00 6 92 9 33 4 00 7 06 10 18 15 4 00 6 90 9 28 4 00 7 03 10 12 20 4 00 6 88 9 23 4 00 7 01 10 06 25 4 00 6 86 9 18 4 01 7 00 10...

Page 15: ...rrect 3 pH electrode SLOPE is off by more than 30 of ideal SLOPE 1 Allow sufficient time for the electrode and ATC Temp probe to stabilize re press SLOPE key to continue the calibration procedure 2 Ch...

Page 16: ...ecognition 90 mV at pH 7 00 or 6 86 pH Electrode SLOPE recognition 30 at pH 4 00 4 01 9 18 and 10 01 Input impedance 1012 Temperature sensor Thermistor 10 k at 25 C Power 9Volt battery Battery Life 10...

Page 17: ...product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis RETURN OF ITEMS Authorization must be obtained from one of our representatives befor...

Page 18: ...coi com Website www jencoi com Jenco Electronics Inc 1F NO 11 Lane 370 Sec 6 Zhongxiao E Rd Nangang Dist Taipei Taiwan TEL 886 2 2782 3226 FAX 886 2 2782 3234 Shanghai Jenco Instruments Ltd 18 Wang Do...

Reviews: