Page 13
14G. Record:
Toggles Record to on and off. When on, the meter stores all readings internally. When
off the meter stops saving readings to the internal memory. Recorded measurements
generate a new file each time the record start button is touched. The format for naming
recorded files is day-mon-year. Record creates a five column .csv file with Time,
Instantaneous, min, max and average results.
14H. Capture
:
Records a single reading at the time the Capture button is pressed. Captured
measurements are appended to the single sample file, in the format date/time,
instantaneous value.
14
I
. Home:
Returns to the home screen.
15 Settings:
The settings icons open the settings display which allows for user customization.
Settings can also be accessed by clicking on the setting drop down menu in the
display screens.
15A. Display Brightness
:
Brightness selection ranges from 1-Dim with lowest battery consumption, to 10-Brightest
with highest battery consumption.
Radiometer Settings
15B. Calibration factors:
15B1. Meter Factor:
Selects calibration factors from the pre-programmed/available list ranging from 1
–
20. Factor selection allows readings in calibrated units such as W, W/cm2, Lux etc.
15B2. Smart Detector Factor:
Smart detectors can also store up to 20 calibration factors. In order to use, or create
cal factors that are saved in the sensor connector, Use Smart Detector Factor must
be enabled, in 15B3.
15B3. USE Smart Detector Factor: ON/OFF
ON enables use of the sensor
’s cal factors. Off only allows customer to use factors
saved in the JEL-2400 meter.