![JDS Uniphase ORL-55 Series Operating Manual Download Page 57](http://html1.mh-extra.com/html/jds-uniphase/orl-55-series/orl-55-series_operating-manual_3632969057.webp)
JDSU
ORL-55
53
8 R
EMOTE
C
ONTROL
Utility commands
:MEM:GROU:MEAS?
:MEM:ID:MEAS?
:MEM:STOR:MEAS
:MEM:REC:MEAS?
:MEM:REC:ALL?
:MEM:DEL:MEAS
:MEM:DEL:GROU
:MEM:DEL:ALL
Command
string
Parameter type / Response type /
Unit / Info
*IDN?
Returns the unique identification of
the device.
Response type:
<STRING_RESPONSE_DATA>
e.g. JDSU Germany GmbH,
ORL-55/01,A-0106,V03.30
*OPC?
Returns “1” as soon as all operations
in progress have been completed.
:SYST
:PERM
:POW
Ensures the device is switched on
permanently i.e. that it does not switch
off after 20 minutes.
:DISP
:CONT
Sets the display contrast.
Parameter type:
<NR1>
Range: 0...15
(0: min, 15: max.)
:DISP
:CONT?
Returns the display contrast.
Range: 0...15
(0: min, 15: max.)