37
5 O
PERATION
ORL MAIN menu
Press
[MENU ENTER]
to open the ORL MAIN menu.
Normalizing
To ensure a high measurement accuracy, it is required to
normalize the unit from time to time. This is done with the
port open, i.e. no DUT is connected.
To normalize the unit:
1. Disconnect DUT from test port.
The normalization can be performed only with an open
test port.
2. Unmount test adapter from test port, clean fibre
surface of test port and mount test adapter.
3. Switch on laser.
4. Press
[MENU ENTER]
to open the ORL MAIN menu.
Normalize
Open Port
Stores the internal noise in ORL
mode as reference value.
Manage
Memory
Display and delete stored data (see
„Memory Management” on page 39
).
Configuration
Change basic settings (see
„Device
overview” on page 9
).
Dangerous laser radiation!
Laser radiation can cause irreparable
damage to the eye and the skin.
!
When the system or test generator is
switched on, never look directly into the
output or into a connected optical fiber.