Digital temperature display 

9.

The digital display of the controller indicates the actual
sample gas heat exchanger temperature.
In normal operation, the status relay alarm is energised.
Above a temperature of 7°c the cooler is on overload.
If the temperature of the heat exchanger rises above
10°c the status relay alarm is de-energised and LED 1
lights up.  (Over temperature alarm)
If the temperature of the sample gas heat exchanger
drops below 0°c the status alarm relays is de-energised.
(Low temperature alarm)
In case of power failure the status alarm relays drops.
The alarm relays is performed as a volt free contact.
Always  operate  contacts  under  specified  ratings.  For
connection of inductive and capacitive loads use suitable
protection circuits (f.i. recovery diodes for inductive and
serial resistance for capacitive loads). Relays are illus-
trated in current- less conditions (fail safe).

The status signal is transferred with a control wire with
open wire ends out of the unit.

Fault display

9.1.

Manual JcS-100

Digitale Temperaturanzeige

9.

Auf dem Display der Regelelektronik wird die aktuelle
Temperatur des Gaswärmetauschers digital angezeigt.
Bei  ordnungsgemäßem  Betrieb  ist  das  Statusrelais
Alarm ist angezogen. Über einer Temperatur von 7°c
liegt eine Überlastung des Messgaskühlers vor. 
Steigt  die  Temperatur  des  Gaswärmetauschers  über
10°c fällt das Statusrelais Alarm ab und LED 1 leuchtet.
(Übertemperaturalarm).
Bei einem Unterschreiten der Temperatur des Gaswär-
metauschers unter 0°c fällt das Statusrelais Alarm. (Un-
tertemperaturalarm)
Bei Stromausfall fällt das Statusrelais Alarm ab. Die kon-
takte des Alarmrelais sind als potentialfreier Wechsler
ausgeführt. Die kontakte sind zu jeder Zeit innerhalb der
spezifizierten Werte zu betreiben. Induktive und kapazi-
tive Lasten sind mit entsprechenden Schutzmaßnahmen
anzuschließen (z.B. Freilaufdioden bei induktiven Lasten
und Serienwiderstände bei kapazitiven Lasten) Relais
sind in stromlosen Zustand (Fail safe) dargestellt.

Das kontaktsignal wird über eine Steuerleitung  mit of-
fenen Enden aus dem Gerät geführt. 

Fehleranzeigen

9.1.

BA_DE_JcS-100_v1.1            

–––––––––––––––––––––––––– [ 16 ] ––––––––––––––––––––––––––

Temperatur Tp > +10°c (LED 1)

1

Temperature Tp > +10°c  (LED 1)

Taste AUF

2

Button UP

Taste AB

3

Button DOWN

Taste Anzeige Temperatur / Eingabe

4

Diplay temperature / Enter button

keine Funktion

5

No function

Ursache 

Anzeige

Display

cause

Temperatursensor Fühlerkurzschluss

F IL

Temperature sensor shorted

Temperatursensor Fühlerbruch

F IH

Temperature sensor break

Programmspeicherfehler

EP

Memory fault

Summary of Contents for JCS-100

Page 1: ...JCT ANALySENTEchNIk GMBh BEDIENUNGSANLEITUNG OPERATING MANUAL 06 16 JCS 100...

Page 2: ...uction 3 1 housing options 3 1 1 Indended use 3 1 2 Versatile usage 3 1 3 General safety information 3 1 4 Service and security 4 1 5 Special instructions 4 1 6 Transport and storage 5 1 7 Technical d...

Page 3: ...ung und Re paratur am Messgask hler auch dann wenn in dieser Bedienungs und Wartungsanleitung nicht ausdr cklich darauf hingewiesen wird Die JcT Analysentechnik Gmbh haftet nicht bei eigen m chtigen n...

Page 4: ...trieb des Ger tes nicht mehr m glich ist muss das Ger t au er Betrieb gesetzt und gegen unbeabsichtigten Betrieb ge sichert werden Es ist anzunehmen dass ein gefahrloser Betrieb nicht mehr m glich ist...

Page 5: ...by the end of its service life Transport und Lagerung 1 7 ACHTUNG Messgask hler d rfen nur in Gebrauchslage transpor tiert werden Es sind die Markierungen auf der Trans portverpackung zu beachten Der...

Page 6: ...kondensatausgang ohne kondensatpumpe 1 4 NPTi oder 3 8 NPTi Schutzart IP 20 EN 60529 Zulassungen Zeichen cE Description sample gas compressor coolers cooling principle compressor cooling Number of gas...

Page 7: ...otentialfreier kontakt max 230 VAc 6 A min 5 VADc 5 mA Anschlussklemmen klemmbereich Federzugklemmen 0 5 mm2 2 5 mm2 BA_DE_JcS 100_v1 1 7 Order Codes JCS 100 series Anzahl der Gaswege Number of gas pa...

Page 8: ...wortliche Spedition und der h ndler zu benachrichtigen Es ist zu berpr fen ob die Ger telieferung Ihrer Bestel lung entspricht Art Nr Verschlei teile 12 90392 Schlauchset kondensatpumpe 5 Stk K1233009...

Page 9: ...orschriften 4 1 Bei Arbeiten am elektrischen Teil des Ger tes ist es vom Netz zu trennen Es ist ausreichender Freiraum um das Ger t f r die Wartung sowie f r die Be und Entl ftung der Mess gasaufberei...

Page 10: ...Anschluss an die Netzversorgung erfolgt ber das schwarze Netzanschlusskabel Das Status Alarmsignal wird ber das graue kabel an geschlossen Entsprechende Nummern sind auf die ka beladern aufgedruckt F...

Page 11: ...ual JcS 100 Inbetriebnahme 5 ACHTUNG Vor dem Einschalten ist sicherzustellen dass die am Ger t eingestellte Betriebsspannung und die Netzspan nung bereinstimmen 1 Dichtheitspr fung der Verbindungen du...

Page 12: ...lare sind mit einer g ltigen Return Aut horisation Number RAN auf Anfrage bei der Serviceabteilung von JcT erh ltlich und stehen auf der JcT Website zum Download bereit JcS 100 Messgask hler sind f r...

Page 13: ...Schritt 3 Step 3 Verschlussriegel im Uhrzeigersinn ff nen und Laufband mit Schlauchein satz entfernen Remove treadmill including the pump tube by rotation of lock clip clockwise Detailansicht Abgenomm...

Page 14: ...the rail until they snap in Schritt 6 Step 6 Verschlussriegel gegen den Uhrzei gersinn drehen bis dieser einrastet Twist back the lock clip counter clock wise until both ends snap in properly Schritt...

Page 15: ...e Replace fan by qualified staff call JcT service Temperature below 0 c Power regulator defective call JcT service Ambient temperature too low Wet sample gas compressor defective call JcT service Samp...

Page 16: ...Bei ordnungsgem em Betrieb ist das Statusrelais Alarm ist angezogen ber einer Temperatur von 7 c liegt eine berlastung des Messgask hlers vor Steigt die Temperatur des Gasw rmetauschers ber 10 c f llt...

Page 17: ...rant dryer Expansionsventil 4 Expansion valve Verdampfer 5 Evaporator W rmetauscher 6 heat exchanger Temperatursensor 7 Temperature sensor kondensatpumpe 8 8 1 condensate pump W rmetauschereinsatz 9 9...

Page 18: ...Dimensions 11 All dimensions in mm Manual JcS 100 Abmessungen 11 Alle Abma e in mm BA_DE_JcS 100_v1 1 18...

Page 19: ...al JcS 100 Wartungsprotokoll 12 BA_DE_JcS 100_v1 1 19 T tigkeit Activity durchgef hrt von performed by Datum Date Unterschrift Signature Pumpenschl uche der kondensat pumpe tauschen Replacement of pum...

Page 20: ...heisenberg Str 4 A 2700 Wr Neustadt Tel 43 0 2622 87 201 Fax 43 0 2622 87 2011 sales jct at Firmenbuchgericht kreisgericht Wr Neustadt FN 119682 d UID Nummer ATU 20559900 Manual JcS 100 BA_DE_JcS 100_...

Reviews: