P/N 960-000180R_Rev. 1 {EDP #233427}
© 2016, JAPAN CASH MACHINE CO., LTD.
6 - 7
Calibration and Testing
DBV® Series DBV-400 Banknote Validator
Section 6
Calibration Preparation
Perform the following steps to prepare the DBV-
400 for Sensor Calibration:
1. Remove electrical power from the DBV-400 Unit.
2. Remove the Cash Box from the DBV-400 Unit.
3. Connect the USB Cable to the USB Connector for
maintenance (Refer to Figure 6-1 and Figure 6-2
for the Tool Requirements and Harness Connector
locations).
4. Set DIP Switch #8 to
ON
5. Apply electrical power to the DBV-400 Unit. The
Bezel LED will flash a
Green
color rate.
6. Launch
the
“
JCM Tool Suite Standard Edition
”
Application (Figure 6-31), then click the “
Service
Mode
” Pull-Down Menu, and select “
Sensor
Adjustment
a
).
7. Confirm that the Sensor Calibration Program
Screen appears (Figure 6-32).
Sensor Calibration
To calibrate the DBV-400 sensors, proceed as
follows:
1. Click the “
Sensor Calibration
a
).
2. Confirm that the “
DBV-400 SENSOR
CALIBRATION
” Screen appears (Figure 6-34).
3. Make sure the Validation Guide Open/Close
Latch (refer to Figure 1-4
n
) is Closed (securely
latched in the Down position on both the left and
right sides). Then click the “
Start
” button
a
) to begin the non-paper calibration.
4. Confirm that the non-paper calibration was
completed with a check mark (Figure 6-35
a
)
appearing next to the “
Calibration without paper
”
text line.
Figure 6-30
DIP SW1 Setting
ON
1
2
3
4
5
6
7
8
Figure 6-31
Launching JCM Tool Suite Stan-
dard Edition/Sensor Adjustment
Figure 6-31
Launching JCM Tool Suite Standard
Edition/Sensor Adjustment Selection
a
Figure 6-32
Figure 6-32
Sensor Calibration Screen 1
Figure 6-33
Figure 6-33
Sensor Calibration Screen 2
a
NOTE: Make sure there are no
foreign objects in the Transport Path.
Figure 6-34
Figure 6-34
Non-Paper Calibration Screen
a
Figure 6-35
Non-Paper Calibration Completion
a