background image

18

Excessive power not pulled. Slowly advance input power (Don’t allow +15 supply to go
higher than 16V). If power pulled is greater that 20W, before reaching 120VAC in, refer
to “Excessive Line Power Test”.

As you ramp out the input voltage, verify that +15V supply stays in regulation. If supply is
non-existent or excessive refer to “+/– 15V Supply Test”. Verify –15V supply is proper, if
not, refer to above test.

Check –V1 (–39V) and +V1 (39V). If non-existent refer to “+/– 39V Supply Test”.

Check –V2 (–77V) and +V2 (39V). If low refer to “+/– 77V Supply Test”.

Verify that “Mute” test point is close to –15V. Remove power to unit and verify mute line
immediately (200mS) goes high (+15). Re-power board and verify mute line initially goes
high and then goes low about 2 seconds later. If any of the above is not true refer to “Ampli-
fier Muting Test”.

C) Confirmation of Low Frequency Chain:
Measure across emitters of Q1 and Q13, for 4mVDC +/– 300uV (20mA). If required adjust
bias pot, VR1. Measure at low frequency amp output (+LF test point) and confirm less than
+/– 50mV of offset. Turn on Generator A and adjust frequency to 400Hz. Adjust Oscillator
amplitude to obtain 60V P/P at amp output. The waveform should be a nice clean sine
wave, with perhaps, a small hint of clipping. The amp will begin to clip at around 62V P/P.
If the amp prematurely clips, verify input to the amp is not clipped. The amp input can be
measured at the junction of R22 and C18. With 60V P/P at the amp output there should be
around 1.9V p/P at the input to the amp. If one or more of the above tests fails. proceed to
“Low Frequency Power Amp Test”.

Adjust oscillator for 1.9V p/P at pin 1 of U2. Verify voltage at input to the filter, pin 1 of J3,
is about 3.5V P/P. If not, proceed to “Low Frequency Filter Test”.

D) Confirmation of High Frequency Chain:
Turn off Generator A. Adjust “High Frequency Level” control, VR201, on signal board, fully
clockwise. Measure at high frequency amp output (+HF test point) and confirm less than
+/– 50mV of offset. Turn on Generator A and adjust frequency to 3kHz. Adjust Oscillator
amplitude to obtain 60V P/P at amp output. The waveform should be a nice clean sine
wave, with perhaps, a small hint of clipping. The amp will begin to clip at around 63V P/P.
Adjust “High Frequency Level” fully counter clockwise and verify that output  level reduces
to around 35V P/P (5dB). Re-adjust “High Frequency Level” fully clockwise. If the amp
prematurely clips, verify input to amp is not clipped. The amp input can be measured at the
junction of R50, R51 and R39. With 60V P/P at the amp output there should be around
2.8V P/P at the input to the amp. If one or more of the above tests fails proceed to “High
Frequency Power Amp Test”.

Adjust oscillator for 3.2V P/P at pin 7 of U2. Verify voltage at the input to the filter, pin 1 of
J3, is about 4V P/P. If not, proceed to “High Frequency Filter Test”.

Summary of Contents for EON15

Page 1: ...oduct repaired with out JBL specified components may not meet the factory specifications The product war ranty may be voided by a repair using unapproved JBL components Test Equipment 1 Audio Precisio...

Page 2: ...t is available for the M115 8 woofer Entire driver should be replaced 124 67001 00X NOTE To recone the EON woofers it is not necessary to disassemble the enclosure The woofer can be reconed from the f...

Page 3: ...ompliance land and spider land with MEK or acetone 7 Remove tape from gap and clean magnet gap with masking tape and with MEK and a rag 8 Unpack recone kit and inspect for damage 9 Verify that the voi...

Page 4: ...est Station 7 Air Gun and Safety Glasses Procedure 1 Record Model number Serial number Date code and compliant for warranty information 2 Lay the enclosure face down on a soft clean surface to avoid d...

Page 5: ...00Hz If no buzzes are detected then test at 2V from 1 500 Hz to 20 kHz for full range operation and audio clarity It will be necessary to lift the baffle to expose the horn so listening tests can be p...

Page 6: ...01 Verify resistance from pin 2 to pin 1 of U202 pin 3 of U202 to ground 10K If not check for opens and shorts around U202 5 Rotate Volume control counter clockwise while monitoring pin 1 of P204 Volt...

Page 7: ...vel for 20V P P on pin 1 of P204 Verify that Peak LED D203 as well as signal LED D201 are on Turn volume control down and verify that Peak indicator turns off A Initial Setup The easiest way to drive...

Page 8: ...to the amp is not clipped The amp input can be measured at the junction of R22 and C18 With 60V P P at the amp output there should be around 1 9V p P at the input to the amp If one or more of the abo...

Page 9: ...ncy Power Amp Test F Supply Test 15V Supply Test Slowly ramp up Variac never allowing the supplies to exceed 16V Measure 15V and 15V supplies Both should be the correct value 5V If a supply is very cl...

Page 10: ...to zero check for 1 collector to emitter short on Q17 2 incorrect value R49 3 shorted TS1 Thermostat should be open when cold and will close at 80 degrees C 4 reversed or shorted Z2 If voltage ramps u...

Page 11: ...ny greater than 2 5VDC Measure resistance of R9 and R21 the correct value is 243 ohms If R9 or R21 value incorrect or if the voltage across these parts is excessive proceed to Voltage Amp Test If R9 a...

Page 12: ...ssover from positive drive to negative drive This actually takes a small chunk sometimes only visible in distortion waveform out of the center of the output waveform This also causes some harmonic dis...

Page 13: ...he characteristics of the waveform and then probe on both sides of C12 C12 is across the input to the current amp If the waveform visibly appears almost identical Same P P voltage same waveform shape...

Page 14: ...s the output bias current and keeps it constant with temperature a Excessive power pulled from AC line This is due to excessive voltage across reference The range measure across C12 should be 1 7V to...

Page 15: ...rify that voltage across R5 equals 870mV 20 If incorrect check Q4 for open or shorted elements Verify that the voltage at the emitter of Q3 is 6V If incorrect check Q3 and Q4 for shorts and opens Veri...

Page 16: ...unction of the first filter section R63 R64 and C37 when this section is operating properly Sweep the oscillator frequency and note if the readings are reasonably close to the values indicated in the...

Page 17: ...n of the filter The second section is a moderate Q Q equals about 2 high pass filter at 55Hz Because of the moderate Q the filter actually peaks The 55Hz Peak Verify the peak frequency 55Hz and the pe...

Page 18: ...esponse at the output of the third section of the filter The third section is a low Q Q equals about 7 high pass filter at 33Hz Verify the peak frequency 55Hz and the peak voltage 5 1V P P Also verify...

Page 19: ...urth section is a moderate Q Q is a little more than 1 low pass filter at 1 4Hz Verify voltages at 100Hz 1kHz 2kHz and 4kHz If the response is incorrect check values of R53 R55 C28 and C33 Reattach li...

Page 20: ...tial If not close check for shorts surrounding pin 10 If no shorts are found replace U1 Verify that pin 9 reflects the output offset polarity The offset should be much greater than 1V If not look for...

Page 21: ...pin 1 of J3 The curve below this line represents the transfer function of the first filter section R61 R71 and C44 when this section is operating properly Sweep the oscillator frequency and note if th...

Page 22: ...curve represents the response at the output of the second section of the filter The second section is a moderate Q Q equals about 2 high pass filter at 1 9kHz Be cause of the higher Q the filter actua...

Page 23: ...High Frequency Response EON Power15 and EON15P 1 230 33...

Page 24: ...Low Frequency Response EON Power15 and EON15P 1 230 34...

Page 25: ...from pin 2 to pin 1 of U2 pin 3 of U2 to ground 3 2k if not check for opened and or shorted traces around U2 5 Rotate volume control counter clockwise while monitoring voltage at the junction of pin...

Page 26: ...Interconnect both boards using a 6 pin cable assembly Advance Volume control on input board R16 fully clockwise Attach power transformer to J5 Adjust bias control R17 fully counter clockwise if at al...

Page 27: ...at the input to the filter pin 1 of J3 is about 2 7V P P If not check for shorts or open circuits around filter IC s U2A U2B and U1A E Excessive Line Power Test Verify the correct polarity and that no...

Page 28: ...or possible shorts in the areas where these parts are All the above parts form the Current Amplifier Providing current gain Everything from Q10 Q12 and before form the Voltage Amplifier Providing volt...

Page 29: ...current amp Q3 Q1 Q7 and Q8 the bias reference needs to have around 2 4V 4 drops across it to turn on all four transistors in the current amplifier The range of adjustment is very small Between 1 7V...

Page 30: ...Low Frequency Response EON PowerSub 40...

Page 31: ...400Hz 2 Verify that voltage at output pin 1 of J1 800mV P P 100mV 3 Verify presence of 400mV P P 50mV at pin 1 of U1 4 Remove oscillator connection to XLR input Verify resistance from pin 2 of J2 to p...

Page 32: ...a 7 pin cable assembly Place the input board in Line mode SW1 out Advance volume control on input board R25 fully clockwise Attach power transformer to J7 Connect the low frequency output to a 4 ohm l...

Page 33: ...ne or more of the above tests fails proceed to High and Low Frequency Power Amp Test If the oscillator is adjusted for 4 9V P P at pin 7 of U4 the voltage at the input to the filter pin 1 of J6 should...

Page 34: ...s the power consumption should go to normal idle level less than 40W if the load is removed If it does not check for shorting from the output pin pin 3 to either the 31V or 31V supplies If not shorted...

Page 35: ...High Frequency Response EON Power10 45...

Page 36: ...Low Frequency Response EON Power10 46...

Page 37: ...nnel 1 Do not plug anything to the Patch nor to the Phone Jacks Adjust volume control R47 Master and R48 Channel 1 fully clockwise and all EQ controls flat at center detented point Verify that the vol...

Page 38: ...he output at pin 1 of J1 is approximately 9 3 dBu Turn High Mid Frequency EQ control fully clockwise and verify that pin 1 of J1 is now approximately 21 2 dBu Turn same knob fully counter clockwise an...

Reviews: