24
25
MVE Changing the pump filters
- Clean the casing with a damp cloth. Make sure to use a soft cloth when cleaning the front.
- Periodically check all cable and tube connections.
- Keep filters clean to ensure proper solder suction and replace them when necessary.
Desoldering process
Use a tip with a larger diameter than the pad to achieve maximum aspiration and thermal efficiency.
Place the tip with the
component terminal in the
hole.
After pressing the desoldering key there is a slight delay until the self-contained vacuum pump
stops. This makes sure that the vacuum circuit is completely empty. If any solder remains are left
on a terminal after desoldering it, resolder it with fresh solder and repeat the desoldering operation.
When the solder liquefies,
gently rotate the tip so that
the compo nent terminal can
be lifted off.
Press the vacuum pump
button long enough to remove
the solder.
3. Aspirating
1. Placing
2. Rotating
The intake tube should be periodically cleaned by the largest rod.
DR560 Tip Care
Important
DO NOT press the vacuum pump button while tinning the desoldering tip, as the fumes given
off by the flux would quickly block the ducts and the air filter.
1.
Select the module from the list of
peripheral connections. Remember your first
connection is denoted as “a”, the second
being “b”, etc. (e.g. MVE_a, MVE_b,...)
2.
Press Menu or Back to save changes.
Once set up, you can change the module
settings by entering the
Peripherals
Menu.
MVE Initial Setup
Peripherals
After connecting the electric desoldering module (MVE-A), enter the
Peripherals
Menu
and select the port which you want to join with the module.
MVE Changing the pump filters
Important
Do not use sharp
pointed objects
to open the
suction filter.
Spare filters
Ref. 0005966
Cotton filters
Ref. 0781046
Suction filter
Ref. 0821830
O Ring
Ref. 0007717
Filter cover
Ref. 0004710
Escape filter
Ref. 0008446
Peripherals
Port 2-DR
17:14
Pedal
None
Module
None
Module
MVE_a
None
www.jbctools.com