background image

124610 Rev. 

B

3

E N G L I S H

I NS TA LL A TIO N

NOTICE: 

Please reference the JAY Fusion Cushion Owner’s
Manual for set-up and adjustment of the cushion. The

 

following instructions pertain to the installation and

 

adjustmen

t of the Air Insert only.

 

A.

INSTALLATION OF THE INFLATION TUBE(S)

Some configurations of the JAY Fusion Cushion with the 

Air 

Insert 

come with the inflation tube(s) un-installed. If this is the 

case, you will have received the following

 

parts:

A second tube will be included
if the air insert was a dual
chamber insert.  A coupling
connector (A) will have already
been installed in the truncated
inflation tubes.

Figure 1 shows a Dual chambered 

Air Insert

 ready to have

the tube length attached by following the steps below.

To install the inflation tube(s)

1. Install the 

Air Insert

into the JAY Fusion foam
base with inner cover as
outlined in the instruc-
tions below.

2. Measure the length of

inflation tube required
for your specific cushion
size (Figure 2). Do this by threading  the tube through the
loops on the inner cover to the front edge.  Measure about
1/2” (13mm) from the front edge of the cushion, and run it
the distance (L) until the open end matches up with the tube
valve coming out of the Insert.

3. Cut the length of inflation tube as measured using a sharp

pair of scissors.

4. Insert the coupling connector into the cut length of each infla-

tion tube,

6. Insert the valve

through the slot in
the outer cover
behind the JAY
Fusion logo.  The
valve is intended to
protrude through
the slot and be
secured by the side
band on the JAY
Fusion  cover. (Figure 6)

CUT

CUT

13mm

13mm

L

L

6

2.

With  the 

Air Insert

  installed,  the

 

inflation  tube(s)  need  to

be secured  to  the  inner  cover.

 

Your 

Air Insert may have one

or  two

 inflation tubes

depend

ing if it is a

single

 chamber insert

or  a

 dual chamber

insert.

 Use the fabric

loops  on

 either side of

the  inner

 cover to

route  the

 inflation

tube(s).
(Figures  4  and  5)

3.

Thread  the  inflation  tube  through  the  loops  starting  at  the
rear  of  the  inner  cover    and  inserting  the  tube  through
each

 

loop,  working  your

 

 

way  around  to  the

front  of  the  product.

4.

If  the 

Air Insert is dual

chambered,  do

 the

same  for  the  sec

ond

inflation  tube  on
the  opposite  side.

5.

Re-install  the  outer  cover  by  sliding  it  over  the  base  and
inner  cover  assembly.    It  is  best  to  hold  the  inflation  valve
in

 

place  with  your  hand  as  you  slide  the  cover  over  the

assem-bly.

1. Air Insert.
2.

1 or 2 lengths of tube with one end terminated by an infla-
tion valve, the other end

 

1

open.

2

A

B.

I

NSTALLATION OF THE AIR INSERT INTO CUSHION

1.

With  the  outer  cover  removed  from  the  JAY  Fusion
foam

 

base,  install  the 

Air Insert

  into  the

 

Well.    Ensure

that  the

 

Air Insert 

is aligned with  the  well

 and that the

inflation tubes

 

are  to  the  rear

 of the

 

cushion.  (See

 Figure 3).

Ensure  the

 Velcro

attached  to  the

 Air

Insert is aligned with

the

 matching Velcro

adhered  to  the  inner
cover.

4

5

3

Reviews: