30
Fig. 4.13 Setting wavelength (CD scale adjustment)
(2) Turn the
″
Scale Correction CD
″
knob located on the rear panel of the amplifier unit
(Fig. 3.4) until the CD value comes within the reference range.
(3) Measure the CD spectrum to check the CD scale.
4.5 Test Signal Check
The test signal is used to check if the electrical system is operating normally. A test
signal can be generated by setting the photometric mode at
″
Test Signal
″
.
<Procedure>
(1) Start up the [Data Monitor] program.
(2) Select the [Parameter...] and [Data mode...] commands from the [Setting] menu
and designate the measurement parameters as shown in Fig. 4.14.
Fig. 4.14 Setting measurement parameters (Test signal check)
(3) Verify that the measurement value of Channel 1 falls within 18 (
±
10%) mdeg.
(4) After that, set the
″
Photometric mode
″
at
″
CD
″
.