27
4.3 Wavelength Accuracy Check and Adjustment
Note
: Before checking wavelength accuracy, warm up the instrument for about one
hour after turning the light source ON.
Note
: Use neodymium glass as the sample.
<Procedure>
(1) Start up the Spectrum Measurement program.
(2) Select the [Parameter...] command from the [Measurement] menu, and designate
the measurement parameters shown in Fig. 4.8.
Fig. 4.8 Measurement parameters (wavelength accuracy check)
(3) Mount neodymium glass in the sample chamber.
(4) Select the [Start] command from the [Measurement] menu in order to perform
measurement.
(5) Using the spectrum analysis program, verify that the peak wavelength of Channel
2 (Fig. 4.9) of the measurement data is 586
±
0.8 nm. If the peak wavelength
does not fall within this range, adjust the instrument according to the following
procedure.
0
1 0 0 0
H T
5 6 0
6 1 0
W a v e l e n g t h [ n m ]
586±0.8nm
Fig. 4.9 HT data of neodymium glass
<Adjustment procedure>
Make adjustment using the adjustment screws on the wavelength lever located at the
bottom of the main unit.
(1) If the main unit is mounted in the optional cabinet, the wavelength cam and