6-10
6.5 Time Program Operation
6.5.1
Running a time program
Start the program from the program monitor screen.
(1)
Starting the time program:
Press the [PRGM RUN] key after the baseline stabilizes. The time program
will start. The [PRGM RUN] key lamp will light, showing that a time program is
currently in progress.
(2)
Stopping the time program:
Press the [PRGM RUN] key while a program is in progress. The [PRGM RUN]
key lamp be turned off and the time program will stop.
6.5.2
Changing parameters while a program is running
The settings for sensitivity, wavelength, response speed, and lamp off timer
can be changed while a program is running. These changes can be made
from the display screens designated by boxes drawn with double lines in Fig.
6.9. The pre-amplified output can also be displayed.
The settings changes are valid until the program progresses and the next
group of settings are activated.
[MONIT]
LAMP OFF TIMER
2.0 H
CD SIG:0.0000 V
UV SIG:1.2740 V
[SHIFT] [ 2]
[SHIFT] [1]
[MONIT]
Lamp off timer screen
Preamp output screen
CD・AU monitor screen
Time・Wavelength monitor screen
PRG 5 T: 0.0
W.L:291 RSPN: SLOW
PRG 5 291 X10
0.000 0.000 STD
Figure 6.9 Changing Parameters While Program is Running
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com