Xonic 5L
22/23
(8) Total Flow Reset function
- In FACTORY MODE menu, enter number
"16"
on PC keyboard and press Enter.
- Total value changes to
“
0.000
”
.
(9) Set Dead-Zone Range adjust
- In FACTORY MODE menu, enter number
"17"
on PC keyboard and press Enter.
- User can adjust starting flow Q0.
- 80mm Q0 operates when inputting a value of 0.8
㎥
/h(3600ps) or more (0.1 m³/h is 456ps)
- 100mm Q0 operates when inputting a value of 1.4
㎥
/h(4800ps) or more (0.1 m³/h is 342ps)
- 150mm Q0 operates when inputting a value of 2.2
㎥
/h(5000ps) or more (0.1 m³/h is 228ps)
- 200mm Q0 operates when inputting a value of 3.52
㎥
/h(6500ps) or more (0.1 m³/h is 182ps)