Xonic 5L
20/23
(5) Reverse KC Factor correction
- In FACTORY MODE menu, enter number
"13"
on PC keyboard and press Enter.
- Enter error value of measured Q3 flow.
- Ex : If error of measured Q3 flow is -5.5%, enter it as a number
-5.5% * 10 =
”5
5
”
(6) Forward DeltaT Offset correction
- In FACTORY MODE menu, enter number
"14"
on PC keyboard and press Enter.
- Enter error value of measured Q2 flow.
- When 80mm measurement error is ±0.1 : Enter value ±12ps / When 80mm measurement
error is ±2.5 : Enter value ±300ps
- When 100mm measurement error is ±0.1 : Enter value ±15ps / When 100mm measurement
error is ±2.5 : Enter value ±375ps
- When 150mm measurement error is ±0.1 : Enter value ±18ps / When 150mm measurement
error is ±2.5 : Enter value ±450ps
- When 200mm measurement error is ±0.1 : Enter value ±22ps / When 200mm measurement
error is ±2.5 : Enter value ±550ps