10. FACTORY MODE    RS232 serial communication   

 

(1) FACTORY MODE menu 

- Menu from 1 to 6 are same as USER MODE menu. 

 

 

(2) Find Dac Level setting 

- In FACTORY MODE menu, enter number 

"10"

 on PC keyboard and press Enter.   

- System will reboot and search Wave signal again. 

- Actual Zero value save in memory does not change. 

         

 

 

 

Summary of Contents for Xonic 5L

Page 1: ...Xonic 5L 1 23 ULTRASONIC WATER METER User s Manual MODEL XONIC 5L ...

Page 2: ...e of Application 4 Xonic 5L Main Board in output terminal 5 Lead Switch operation function 6 LCD Display 7 PC and Xonic 5L communication cable connection 8 PC USB TO RS232 serial communication setting 9 USER MODE RS232 serial communication 10 FACTORY MODE RS232 serial communication ...

Page 3: ...quipped with lithium primary battery that can operate up to 8 years General specifications are as in below Principle Ultrasonic Transit Time Velocity 0 02 10m sec Flange Material Stainless Steel 304 Sensor Type Insertion Power Supply 3 6V Battery Life 8 years Delivery Items electronics cable battery flange ...

Page 4: ... that has enough straight pipe run 7D for upstream and 3D for downstream 전단 7D 후단 3D Make sure that the sensor part is not located at the top Error may occur due to the air bubble in the upper part Incorrect Installation Correct Installation Correctly installed water meter ...

Page 5: ...Xonic 5L ultrasonic water meter The flow rate can be adjusted to fit 80mm 100mm 150mm 200mm 250mm and 300mm flanges of Xonic 5L ultrasonic flowmeter There are Factory mode and User mode various flow related data can be calibrated or selected 4 Xonic 5L Main Board in output terminal 1 Main Board ...

Page 6: ..._3 6V 2Pin GND 5 CN502_Firmware Write Connector CONNECTOR spec Yeonho Electronics 12505WS 10P 1Pin JTAG_TOD 2Pin JTAG_TOI 3Pin JTAG_TMS 4Pin JTAG_TCK 5Pin RESET 6Pin NC 7Pin 3 6V 8Pin 3 6V 9Pin GND 10Pin GND 6 CN503_External battery input connector Use external battery DC 3 6V CONNECTOR spec Yeonho Electronics 12505WS 4P 1Pin RS232_TX 2Pin RS232_RX 3Pin GND 4Pin External battery_VALID 7 CN505_RS23...

Page 7: ... power OFF After LCD display screen is off bring neodymium magnet into contact with Lead Switch SW102 Turn the power ON The icon and firmware version will be displayed on LCD for about 30 seconds to 1 minute Completed when total flow and m unit are displayed on LCD Once actual zero is completed it starts to operate in Factory mode If no battery sign then it is Factory mode 5 Switch to Factory mode...

Page 8: ...Xonic 5L 8 23 6 LCD Display 1 Full screen with icon and number 2 User mode screen with battery sign 3 Factory mode screen no battery sign ...

Page 9: ...Xonic 5L 9 23 4 Firmware version Actual zero operation screen 6 digit integers indicate the firmware version 3 decimal places indicate the pipe diameter ...

Page 10: ...L BATTERY CIRCULAR 4 USB TO RS232 Serial Converter 5 Connect XONIC 5L and PC comm cable Connect XONIC 5L RS232 CIRCULAR cable A to Main Board CN505 Connect XONIC 5L RS232 CIRCULAR B and XONIC 5 RS232 CABLE C Connect XONIC 5 RS232 CABLE D and USB TO RS232 Serial Converter DB9M Connect USB TO RS232 Serial Converter USB to a PC A B E F C D ...

Page 11: ...Xonic 5L 11 23 8 PC USB TO RS232 serial communication setting 1 PC USB communication setting ...

Page 12: ...er Universal serial bus controller check if the USB TO RS232 Serial Converter is connected to the PC USB terminal If there is no USB Serial Converter in Control Panel Device Manager Universal serial bus controller reinstall the USB TO RS232 Serial Converter operation software on the PC 2 PC Hyper Terminal serial communication setting Port Set by checking USB Serial Port COM number in Port COM LPT ...

Page 13: ...Xonic 5L 13 23 9 USER MODE RS232 serial communication 1 USER MODE menu 2 Display Information In USER MODE menu enter number 1 on PC keyboard and press Enter ...

Page 14: ...digits Ex 01 12 day xx Enter the day and press Enter 2 digits Ex 01 31 day of week x Enter the day of week and press Enter 1 digit Ex Sunday 0 Hour xx Enter the hour and press Enter 2 digits Ex 01 24 Minute xx Enter the minute and press Enter 2 digits Ex 00 59 Second xx Enter the second and press Enter 2 digits Ex 00 59 After completion check year month day day of week and time in RTC TIME Info ...

Page 15: ...ogging Disable Log data is not saved 3 Cancel No selection and returns to main menu 6 Invalid TIME SET error message in Logging On Off Selection In USER MODE menu enter number 3 on PC keyboard and press Enter On PC keyboard enter number 1 and press Enter When error message Invalid TIME SET appears set the Display RTC Time again 7 Logging On Off Selection setting ...

Page 16: ...ve log data Completer when Init FRAM message appears Caution if you select 1 Logging Enable Existing Log Data Loss the existing data will be deleted In USER MODE menu enter number 3 on PC keyboard and press Enter In PC keyboard enter number 2 and press Enter If Current Logging state 2 Logging Disable message appears log data is not saved ...

Page 17: ...rd and press Enter Current Log Interval displays the days that can be saved with the selected Log Interval time 9 Log Data Output setting In USER MODE menu enter number 5 on PC keyboard and press Enter Maximum number of log data is 16383 Caution Save Xonic 5L log data on PC using log function of Hyper Terminal 10 Find DAC OP On Off setting Do not use ...

Page 18: ...on 1 FACTORY MODE menu Menu from 1 to 6 are same as USER MODE menu 2 Find Dac Level setting In FACTORY MODE menu enter number 10 on PC keyboard and press Enter System will reboot and search Wave signal again Actual Zero value save in memory does not change ...

Page 19: ...d enter the value in numbers Ex If water temperature is 25 and sound speed is 1496 7m s enter sound speed as a number 1496 7m s 10 14967 Check temperature in Display Information 4 Forward KC Factor correction In FACTORY MODE menu enter number 12 on PC keyboard and press Enter Enter error value of measured Q3 flow Ex If error of measured Q3 flow is 5 5 enter it as a number 5 5 10 55 ...

Page 20: ... on PC keyboard and press Enter Enter error value of measured Q2 flow When 80mm measurement error is 0 1 Enter value 12ps When 80mm measurement error is 2 5 Enter value 300ps When 100mm measurement error is 0 1 Enter value 15ps When 100mm measurement error is 2 5 Enter value 375ps When 150mm measurement error is 0 1 Enter value 18ps When 150mm measurement error is 2 5 Enter value 450ps When 200mm ...

Page 21: ... Enter value 12ps When 80mm measurement error is 2 5 Enter value 300ps When 100mm measurement error is 0 1 Enter value 15ps When 100mm measurement error is 2 5 Enter value 375ps When 150mm measurement error is 0 1 Enter value 18ps When 150mm measurement error is 2 5 Enter value 450ps When 200mm measurement error is 0 1 Enter value 22ps When 200mm measurement error is 2 5 Enter value 550ps ...

Page 22: ...mber 17 on PC keyboard and press Enter User can adjust starting flow Q0 80mm Q0 operates when inputting a value of 0 8 h 3600ps or more 0 1 m h is 456ps 100mm Q0 operates when inputting a value of 1 4 h 4800ps or more 0 1 m h is 342ps 150mm Q0 operates when inputting a value of 2 2 h 5000ps or more 0 1 m h is 228ps 200mm Q0 operates when inputting a value of 3 52 h 6500ps or more 0 1 m h is 182ps ...

Page 23: ...nd if it is above this range enter continuously Ex If the input value is 80000 enter it 3 times by 30000 11 Actual Zero Value Add adjust In FACTORY MODE menu enter number 19 on PC keyboard and press Enter Actual Zero Value can be inputted Input below 30000 30000 at once Total value can be inputted below 30000 30000 and if it is above this range enter continuously Ex If the input value is 80000 ent...