4.8 Install
–
Install Info
User could review the install information of the selected path in this display.
Caution) The Path Length will show once the program has finished the installation.
4.9 Install –
Actual Zero
User should perform Actual Zero when the flow is actual stopped but the flowmeter
shows other values instead of “0”. This function, Actual Zero can help user to adjust
the Zero Point for flowmeter.
Caution) Use the function when the flow is stopped.
Actual Zero User
Actual Zero User, is the automatic function to do actual zero by the software.
Actual Zero Set
Actual Zero Set, is the manual function to do actual zero by user. Usually, user will
use the function after perform “Actual Zero User/Actual Zero Auto”. To clear the
negative value.
Actual Zero Set
Actual Zero Reset, is the function to clear the data. Usually, user will use the function
after perform “Actual Zero User/Actual Zero Auto”. To clear the negative value.
Actual Zero Auto
Actual Zero Auto, is the automatic function to do actual zero by the software when
user is UNABLE to stop the flow.
4.9 Install –
Sampling Clk
Flow computer will select Sampling Clock automatically, therefore, user does not need
to change.
INSTALL
1 . SENSOR TYPE
2 . PATH HEIGHT
3 . PATH LENGTH
4 . PATH ANGLE
5 . PATH HEIGHT OPT
6 . AUTO INSTALL
7 . RESERVE INSTALL
8 . INSTALL INFO.
9 . ACTUAL ZERO
0 . SAMPLING CLK
INSTALL INFO.
-Sensor:
LTR-L (F 0.3MHz)
-Height: 100mm
-Inst. Path Length Not
Installed
INSTALL
1 . SENSOR TYPE
2 . PATH HEIGHT
3 . PATH LENGTH
4 . PATH ANGLE
5 . PATH HEIGHT OPT
6 . AUTO INSTALL
7 . RESERVE INSTALL
8 . INSTALL INFO.
9 . ACTUAL ZERO
0 . SAMPLING CLK
1 . ACTUAL ZERO USER
2 . ACTUAL ZERO SET
3 . ACTUAL ZERO RESET
4 . ACTUAL ZERO AUTO
ACTUAL ZERO
Press [
MENU
] – 4. INSTALL –
9. ACTUAL ZERO
▪
Actual Zero User (automatic adjustment by software)
▪
Actual Zero Set (manual adjustment by user)
▪
Actual Zero Reset (clear data to zero)
▪
Actual Zero Auto (when user couldn’t stop the flow)
Press [
MENU
] – 4.INSTALL –
9. ACTUAL ZERO –
1.ACTUAL ZERO USER
➢
Press [ENT] to perform the function.
➢
The software will take about 30 second to reset the zero point for flowmeter
➢
After it finished, user will see a negative value in this menu.
Press [
MENU
] – 4. INSTALL –
9. ACTUAL ZERO –
4.ACTUAL ZERO AUTO
➢
Press [ENT] to perform the function.
➢
The software will take about 30 second to reset the zero point for flowmeter.
➢
After it finished, user will see a negative value in this menu.
Press [
MENU
] – 4.INSTALL –
8.INSTALL INFO
Install Path Length
➢
Before running “Auto Install”, it will show
“Not
installed”
even user has input the path length.
➢
The actual path length will be calculated
automatically
after user run “4.6 Auto Install”.
➢
If the software has performed “Reserve Install”, it will
show
“Install Reserved”
.
Press [
MENU
] – 4.INSTALL –
9. ACTUAL ZERO –
2.ACTUAL ZERO SET
➢
Input the zero value with [NUM].
➢
Press [ENT] to save the data.
Press [
MENU
] – 4. INSTALL –
9. ACTUAL ZERO –
3.ACTUAL ZERO RESET
➢
Press [ENT] to clear the data.
Summary of Contents for Xonic 100 Series
Page 43: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 43...
Page 44: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 44...
Page 45: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 45...
Page 46: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 46...
Page 47: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 47...