background image

Xonic

®

  100  Series 

 

Jain  Technology

 

www.jain.co.kr

    Tel.  82-2-856-4114  Fax.  82-2-856-9503 

26 

4.8 Install

 –   

Install Info

 

User could review the install information of the selected path in this display. 

Caution) The Path Length will show once the program has finished the installation.

     

 

 

 

 

4.9 Install –   

Actual Zero 

User should perform Actual Zero when the flow is actual stopped but the flowmeter 
shows other values instead of “0”. This function, Actual Zero can help user to adjust 
the Zero Point for flowmeter. 

Caution) Use the function when the flow is stopped.

 

 

 

 

 

Actual Zero User 

Actual Zero User, is the automatic function to do actual zero by the software. 

 

Actual Zero Set 

Actual Zero Set, is the manual function to do actual zero by user. Usually, user will 
use the function after perform “Actual Zero User/Actual Zero Auto”. To clear the 
negative value.

 

 

Actual  Zero  Set

 

Actual Zero Reset, is the function to clear the data. Usually, user will use the function 
after perform “Actual Zero User/Actual Zero Auto”. To clear the negative value. 

 

Actual Zero Auto 

Actual Zero Auto, is the automatic function to do actual zero by the software when 
user is UNABLE to stop the flow. 

 

 

4.9 Install –   

Sampling Clk 

Flow computer will select Sampling Clock automatically, therefore, user does not need 
to change.   

INSTALL
1 . SENSOR TYPE
2 . PATH HEIGHT
3 . PATH LENGTH
4 . PATH ANGLE
5 . PATH HEIGHT OPT
6 . AUTO INSTALL
7 .  RESERVE INSTALL
8 . INSTALL INFO.
9 . ACTUAL ZERO
0 . SAMPLING CLK

INSTALL INFO.
 -Sensor:

LTR-L (F 0.3MHz)

 -Height: 100mm
 -Inst. Path Length Not

Installed

INSTALL
1 . SENSOR TYPE
2 . PATH HEIGHT
3 . PATH LENGTH
4 . PATH ANGLE
5 . PATH HEIGHT OPT
6 . AUTO INSTALL
7 .  RESERVE INSTALL
8 . INSTALL INFO.
9 . ACTUAL ZERO
0 . SAMPLING CLK

1 . ACTUAL ZERO USER
2 . ACTUAL ZERO SET
3 . ACTUAL ZERO RESET
4 . ACTUAL ZERO AUTO

ACTUAL ZERO

Press [

MENU

] – 4. INSTALL –

 9. ACTUAL ZERO 

 

 

Actual Zero User (automatic adjustment by software) 

 

Actual Zero Set (manual adjustment by user) 

 

Actual Zero Reset (clear data to zero) 

 

Actual Zero Auto (when user couldn’t stop the flow) 

 

Press [

MENU

] – 4.INSTALL –

 

9. ACTUAL ZERO – 

1.ACTUAL ZERO USER 

 

 

Press [ENT] to perform the function. 

 

The software will take about 30 second to reset the zero point for flowmeter 

 

After it finished, user will see a negative value in this menu.   

 

Press [

MENU

] – 4. INSTALL –

 

9. ACTUAL ZERO – 

4.ACTUAL ZERO AUTO 

 

 

Press [ENT] to perform the function. 

 

The software will take about 30 second to reset the zero point for flowmeter.   

 

After it finished, user will see a negative value in this menu.   

Press [

MENU

] – 4.INSTALL –

 8.INSTALL INFO 

 

Install Path Length   

 

Before running “Auto Install”, it will show 

“Not 

installed”

 even user has input the path length. 

 

The actual path length will be calculated 
automatically

 after user run “4.6 Auto Install”.   

 

If the software has performed “Reserve Install”, it will 
show 

“Install Reserved”

.     

Press [

MENU

] – 4.INSTALL –

 

9. ACTUAL ZERO – 

2.ACTUAL ZERO SET 

 

 

Input the zero value with [NUM]. 

 

Press [ENT] to save the data. 

 

Press [

MENU

] – 4. INSTALL –

 

9. ACTUAL ZERO – 

3.ACTUAL ZERO RESET 

 

 

Press [ENT] to clear the data. 

 
 

Summary of Contents for Xonic 100 Series

Page 1: ...Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 Technical Information Xonic 100 Series Open Channel Ultrasonic Flowmeter Open Channel Guidebook Instruction Manual...

Page 2: ...uctions Quick Analog Output 12 Instructions Analog Input for Level Meter 14 Instructions Data Logger 16 Setting Operation General Setting 20 Setting Operation Channel Setup 22 Setting Operation Instal...

Page 3: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 3 Delivery Items Flow Computer Transducer Bracket Cable...

Page 4: ...r Manure Etc Steel Factory and Mining Industries Limestone Slurries Deionized Water Etc Power Plants Irrigation Canal River Stream etc Features Patented AR Mode Anti Round Specifications Principle o A...

Page 5: ...e Name Press to change the alphabet while inputting Press to input minus sign and symbols while inputting Press to enter the choice finish inputting data Press to select the choice in menu Press to in...

Page 6: ...concentrations of suspended solids air or a large noise component Please read this manual carefully before installation e g Open Channel Dual Path Step 1 Fixing of Mounting Bracket Selecting the right...

Page 7: ...is measured from the CENTER of the sensor to the mounting wall In order to receive a good signal for best accuracy all transducers must face diagonally For trapezoidal or circular channels rotate the...

Page 8: ...82 2 856 4114 Fax 82 2 856 9503 8 Step 5 Supply Power to Flow Computer Connect the power cable of the flow computer and turn on the power Step 6 Input the Values on Flow Computer Please refer to next...

Page 9: ...Input Unit Step 2 Select the Path for introducing the Install Parameter 2 Path Select User must select the path first to setup the configuration data in other tabs Step 3 Channel Shape Setup 3 Channe...

Page 10: ...or 2 Path Height the mounting height from the bottom 3 Path Length the length between 2 transducers Lens to Lens MAIN MENU 1 GENERAL 2 PATH SELECT 3 CH SETUP 4 INSTALL P1 5 OPERATE 6 FLOW 7 IN OUTPUT...

Page 11: ...or a better performance 7 Reserve Install click SET for each path to enable the auto installation After setup the information for Path One User needs to go back to 2 Path Select to select other paths...

Page 12: ...ANALOG OUT 1 or 2 ANALOG OUT 2 7 1 1 Output Data Choose the required data for output 1 None Disable 2 Flow 3 Velocity 4 Analog In 1 5 Analog In 2 7 1 2 Output Mode Depending on the measuring applicat...

Page 13: ...nd wiring the connecting devices to the selected OUTPUT on flowmeter s PC board as below wiring diagram AO1 4 20 mA Output AO1 AO2 AO2 AI1 4 20 mA Input AI1 AI2 AI2 Flowmeter Step 3 Power On After reb...

Page 14: ...ption User can write some description for the input data 7 5 7 Check Input Data After connected the input device user can check the input data here Step 2 Level Meter Setup In the case of Open Channel...

Page 15: ...Step 4 Power On Check Input Data After connecting user shall see the Input data in the MAIN display as below Or user can check in the tab 7 5 7 Check Input Data to check the input data as well SITE___...

Page 16: ...INTERNAL MEMORY 5 LOG ENABLE 2 ENABLE Step 3 Download DNW exe Download the software in user s Laptop Caution Please contact the manufacture for the software DNW allows user to 1 Output the logger dat...

Page 17: ...Open the Device Manager on user s Laptop Find PORTS COM LPT and click Communications Port Open the tab Port Settings and click Advanced Feel free to set the COM Port Number to COM 1 COM 2 COM 3 COM 4...

Page 18: ...t this step Run the Software DNW exe and click Configuration User will see the window of UART USB Options as below Set the Baud Rate at 115200 and the COM Port that user just set After return to DNW o...

Page 19: ...MENU 8 DATA LOG 3 MEMORY 1 INTERNAL MEMORY 1 LOG OUTPUT 2 LOG OUTPUT Press ENT to output the data Make sure 2 ENABLE is selected on 5 LOG ENABLE menu 1 TIME SET 2 RS 232C 3 MEMORY 4 WAVE DATA DATALOG...

Page 20: ...5 CHANNEL SHAPE 6 LEVEL METER SET SYSTEM LOCK 1 UNLOCK LOCK 2 CHANGEPASSWORD GENERAL 1 SYSTEM LOCK 2 SITENAME 3 APPLICATION 4 UNIT 5 CHANNEL SHAPE 6 LEVEL METER SET SITENAME SITE_ _ _ _ GENERAL 1 SYS...

Page 21: ...smitter User can use either Analog In 1 or Analog In 2 for input level data If Channel is always full select Set Full Set up the input unit of the level transmitter 1 INPUT 2 UNIT LEVEL METER SET INPU...

Page 22: ...TALL 5 OPERATE 6 FLOW 7 IN OUTPUT 8 DATALOG 9 DIAG PATH SELECT 1 PATH 1 2 PATH 2 3 PATH 3 4 PATH 4 MAIN MENU 1 GENERAL 2 PATH SELECT 3 CH SETUP 4 INSTALL 5 OPERATE 6 FLOW 7 IN OUTPUT 8 DATALOG 9 DIAG...

Page 23: ...OPT 6 AUTO INSTALL 7 RESERVE INSTALL 8 INSTALL INFO 9 ACTUAL ZERO 0 SAMPLING CLK INSTALL 1 SENSOR TYPE 2 PATH HEIGHT 3 PATH LENGTH 4 PATH ANGLE 5 PATH HEIGHT OPT 6 AUTO INSTALL 7 RESERVE INSTALL 8 INS...

Page 24: ...n INSTALL 1 SENSOR TYPE 2 PATH HEIGHT 3 PATH LENGTH 4 PATH ANGLE 5 PATH HEIGHT OPT 6 AUTO INSTALL 7 RESERVE INSTALL 8 INSTALL INFO 9 ACTUAL ZERO 0 SAMPLING CLK 60 000 PATH ANGLE UNIT Degree INSTALL 1...

Page 25: ...150mm And once the level reaches the Expected Height of Path 2 300mm It will perform the installation automatically again for Path 2 How to make sure the installation is completed There are some point...

Page 26: ...SERVE INSTALL 8 INSTALL INFO 9 ACTUAL ZERO 0 SAMPLING CLK INSTALL INFO Sensor LTR L F 0 3MHz Height 100mm Inst Path Length Not Installed INSTALL 1 SENSOR TYPE 2 PATH HEIGHT 3 PATH LENGTH 4 PATH ANGLE...

Page 27: ...of the channel bottom to aviod false echoes from obstructions MAIN MENU 1 GENERAL 2 PATH SELECT 3 CH SETUP 4 INSTALL 1 5 OPERATE 6 FLOW 7 IN OUTPUT 8 DATALOG 9 DIAG 1 OPERATE 1 UPPER FLOW LIMIT 2 LOWE...

Page 28: ...2 LOWER FLOW LIMIT 3 DEAD LEVEL 4 DEAD ZONE 5 FLOW AVERAGE TIME 6 TOTAL FLOW SET 7 ALARM 8 CALIBRATION 9 ENABLE AGC 0 DAMPING sec 5 FLOW AVERAGE TIME OPERATE 1 UPPER FLOW LIMIT 2 LOWER FLOW LIMIT 3 D...

Page 29: ...TI POINT SET 3 Kc SET MULTI POINT SET 1 VIEW 2 ADD 3 DELETE 1 METHOD 2 MULTI POINT SET 3 Kc SET CALIBRATION 1 000 Kc SET Unit None OPERATE 1 UPPER FLOW LIMIT 2 LOWER FLOW LIMIT 3 DEAD LEVEL 4 DEAD ZON...

Page 30: ...E UNIT 3 FLOW RESOLUTION 4 FLOW SCALE 5 TOTAL VOLUME UNIT 6 TOTAL RESOLUTION 7 TOTAL SCALE 8 BATCH TOTAL 9 TOTALIZER MODE 1 SECOND 2 MINUTE 3 HOUR 4 DAY FLOW TIME UNIT 1 FLOW VOLUME UNIT 2 FLOW TIME U...

Page 31: ...FLOW SCALE 5 TOTAL VOLUME UNIT 6 TOTAL RESOLUTION 7 TOTAL SCALE 8 BATCH TOTAL 9 TOTALIZER MODE FLOW 1 xxxx 2 xxx x 3 xx xx 4 x xxx TOTAL RESOLUTION 1 FLOW VOLUME UNIT 2 FLOW TIME UNIT 3 FLOW RESOLUTIO...

Page 32: ...9 DIAG IN OUTPUT 1 ANALOG OUT 1 2 ANALOG OUT 2 3 RELAY OUT 1 4 RELAY OUT 2 5 ANALOG IN 1 6 ANALOG IN 2 ANALOG OUT 1 1 OUTPUT DATA 2 OUTPUT MODE 3 CALIBRATION_MIN 4 CALIBRATION_MAX 5 SPAN MIN 6 SPAN MA...

Page 33: ...5 SPAN MIN 6 SPAN MAX 7 2mA SET ANALOG OUT 1 SPAN MAX m h 0 1 OUTPUT DATA 2 OUTPUT MODE 3 DUTY SET 4 REV RELAY OUT 1 OUTPUT DATA 1 RELAY OFF 2 RELAY ON 3 HIGH FLOW 4 LOW FLOW 5 HIGH LOW FLOW 6 HIGH SP...

Page 34: ...user s input device 1 OUTPUT DATA 2 OUTPUT MODE 3 DUTY SET 4 REV RELAY OUT 1 1 REV00 2 REV04 REV 1 SET ENABLE 2 CALIBRATION_MIN 3 CALIBRATION_MAX 4 MIN INPUT SPAN 5 MAX INPUT SPAN 6 DESCRIPTION 7 CHE...

Page 35: ...tting for data logger 1 SET ENABLE 2 CALIBRATION_MIN 3 CALIBRATION_MAX 4 MIN INPUT SPAN 5 MAX INPUT SPAN 6 DESCRIPTION 7 CHECK INPUT DATA ANALOG IN 1 DESCRIPTION A_IN1 1 SET ENABLE 2 CALIBRATION_MIN 3...

Page 36: ...BITS 5 LINE FEED 6 NETWORK ID CONFIGURE 1 8Bits 2 7Bits DATA BITS 1 BAUD RATE 2 DATA BITS 3 PARITY 4 STOP BITS 5 LINE FEED 6 NETWORK ID CONFIGURE 1 NO PARITY 2 EVEN PARITY 3 ODD PARITY PARITY 1 BAUD...

Page 37: ...G TIME SYNC 7 LOG ENABLE 8 COMM MODE 9 SELECT COMM HFTA H Header S Site name N Channel No D Date C Time F Flow T Total I1 I2 AnalogIn1 2 U Unit A Alarm V Velocity FORMAT RS 232C 1 CONFIGURE 2 HEADER 3...

Page 38: ...INTERVAL 6 LOG TIME SYNC 7 LOG ENABLE 8 COMM MODE 9 SELECT COMM 1 RS232 2 RS485 SELECT COMM 1 LOG OUTPUT 2 FORMAT 3 LOG INTERVAL 4 LOG TIME SYNC 5 LOG ENABLE 6 PORT SELECT 7 MEMORY CLEAR MEMORY LOG O...

Page 39: ...oad the Signal Wave Data MEMORY 1 LOG OUTPUT 2 FORMAT 3 LOG INTERVAL 4 LOG TIME SYNC 5 LOG ENABLE 6 PORT SELECT 7 MEMORY CLEAR LOG ENABLE 1 DISABLE 2 ENABLE MEMORY 1 LOG OUTPUT 2 FORMAT 3 LOG INTERVAL...

Page 40: ...evel is amplitude level of signal The value will be set and calculated automatically by the flowmeter MAIN MENU 1 GENERAL 2 PATH SELECT 3 CH SETUP 4 INSTALL 5 OPERATE 6 FLOW 7 IN OUTPUT 8 DATALOG 9 DI...

Page 41: ...2 RISC 3 PULSE COUNT 4 GAIN LEVEL 5 HOLD FLOW 6 CONTROL HOLD FLOW Unit m h _ _ _ _ CONTROL 1 MAKEUPCONTROL 2 ANTI FAULT TIME 3 MANNING HEIGHT 4 MANNING VELOCITY 1 DISABLE 2 ENABLE MAKEUPCONTROL Press...

Page 42: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 42 DRAWINGS...

Page 43: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 43...

Page 44: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 44...

Page 45: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 45...

Page 46: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 46...

Page 47: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 47...

Reviews: