CM-140 GE / -RA/-UV / CB-140GE/-RA
- 49 -
Trigger
CCD Exposure
350L :10.71ms(GE)
20.615ms(UV )
EEN
FVAL
DVAL
2L
~
3
L
GE:(Full:63μs
~ 94μs,V Binning 89μs ~ 126μs)
UV:(Full:117.8μs ~
176.7µs, V Binning 149.6µ
s ~
224.4µs)
Note: When PE is set at 1052 or the shutter is set at OFF, EEN is always HIGH.
Fig.32 RCT mode timing
10.2.5 Sequential Trigger Mode (Pre-Select trigger)
The ROI, Shutter and Gain values can be preset for up to 10 sequential trigger events. With every
trigger input, the image data from the preset sequence is output as described below.
Trigger
Sequence
Operation
Signals added to trigger can be selected by Trigger Source. The camera will function on the rising
edge of the trigger and Negative or Positive should be determined accordingly.
For the sequence, the following default settings are installed.
ID
ROI
Shutter
Gain
Width Height Offset
X
Offset
Y
1
1392 1040
0
0
1052
0
2
1392 1040
0
0
1052
0
3
1392 1040
0
0
1052
0
4
1392 1040
0
0
1052
0
5
1392 1040
0
0
1052
0
6
1392 1040
0
0
1052
0
7
1392 1040
0
0
1052
0
8
1392 1040
0
0
1052
0
9
1392 1040
0
0
1052
0
10
1392 1040
0
0
1052
0
Sequence 1
Sequence 4
Sequence 3
Sequence 2