CM-141MCL/PMCL/RA / CB-141MCL/PMCL/-RA
28
6.3.5 Smear less EPS Trigger mode
This mode is EPS trigger mode with smear less function. Once the external trigger is input, a
fast dump is activated. Then exposure starts with the preset shutter speed.
This mode can eliminate smear on the upper part of the image.
This mode functions only in LVAL asynchronous mode.
To use this mode:
Set function:
Trigger mode Smear less Edge PreSelect
Scanning
Full, Partial
Vertical binning ON / OFF (CM-141MCL only)
Shutter mode Preset, Programmable, Fine Exposure
Shutter speed 1/60 to 1/10,000
Programmable exposure 2L to 1052 L
Accumulation
LVAL async only
Other functions and settings
Input:
Ext. trigger.
12-pin Hirose or Camera Link
Important notes on using this mode
Trigger pulse >2 LVAL to <
The following table shows minimum trigger interval in synchronous accumulation mode
Full scan
1268 L
2/3 Partial
989 L
1/2 Partial
850 L
1/4 Partial
642 L
1/8 Partial
438 L
1/2 V Binning
744 L
Fig.31 Smearless EPS mode timing
Trigger
CCD
Exposure
EEN
FVAL
DVAL
211L
2L to 3L
Trigger Invalid Period
Smearless Period