JWI-700D/700S 11
always
:
Tare/Zero function acts by pressing down key
TARE
or
ZERO
even if it is not
stable
12.
Power saving function
on:
Enable the power saving function
off:
Disable the power saving function
13.
Initialization
Press
M-/SET,
then press
M+/SAVE
to start initialization. After the window displays
“ok”,
the initialization is finished
6. Calibration
Note: Before calibration, please set the unit first. The unit used in calibration is the one that has
been set before. During the calibration procedure, press
ZERO/ESC
to return to normal
weighing mode without saving.
Here we take 3kg/10g as an example
1. Press and hold
TARE
while powering on. Do not release it till the window displays
“
CAL
”.
2. With no load on the weighing pan, press
TARE
to start zero point calibration.
“
”is
blanking at the left bottom.
3. Wait till the window displays the first calibration value.
“
”appears at the left bottom.
Note: The first calibration value is default. If you need to change the value, do as the following:
Press
M-/SET
to enter the value setting. Press
MR/
or
TARE/
to move the key leftwards
or rightwards. Press
+
/PRINT/HI
or
-/
H
OLD/
LO
to change the value. Press
M+/SAVE
to
save.
4. Put the corresponding weight on the weighing pan, and then press
TARE
to complete the
first point calibration.
“
”appears at the left bottom.
Note: After the first point calibration, the window can display the weight value. If no need for
the other point calibration, move to step 6 to finish the calibration procedure.
5. Add another mass to the current weight. The window will show the total weight. Press
TARE