Jadever 1.5K User Manual Download Page 11

 

 

 

 

4)  Wait  till 

appears,  put  weights

JWL-6K, 2/3 of full load is 4 kg)  

5) Wait till  

  

appears, put weights of full load on the weighing pan (E.g. for JWL

load is 6kg)  

 

6) The calibration procedure ends up with a symbol

7) Remove all the weights, and then press key 

 

  Single Point Calibration 

 

1

 Press and hold key 

M- 

while powering on the scale. The screen displays 

release the  

 

2

 Press key 

TARE 

to enter the zero point

3

 Wait  till  the 

”  appears  (with  the  right

shift leftwards, key 

PRINT/HI 

to shift rightwards, key 

key 

M- 

to decrease setting value.  

4

 Put the corresponding weights on the  weighing pan and press key 

procedure starts. 

5

A few seconds later, the calibration procedure is completed with 

the  screen.  Remove  all  the  weights  and  press  key 

mode. 

7. Setting Mode & Description of Parameter

1) Long press key 

MODE 

for 3 seconds

reverts to available setting items .There

Auxiliary display in the Lower Left corner

2) Press key 

M+ 

or key 

M- 

to step through all the available setting items ,and release on

 

JWL-6K, 1/3 of full load is 2 kg) 

Vetek 

weights  of  2/3  of  full  load  on  the  weighing

 

appears, put weights of full load on the weighing pan (E.g. for JWL

The calibration procedure ends up with a symbol of “ 

Remove all the weights, and then press key 

TARE 

to return to weighing mode.

while powering on the scale. The screen displays 

to enter the zero point calibration 

 

”  appears  (with  the  right-most  digit  blinking),  press  key 

to shift rightwards, key 

M+   

to increase setting values and 

value.  

 

Put  the corresponding weights on  the  weighing  pan and press key 

N/G

the calibration procedure is completed with 

” displayed 

the  screen.  Remove  all  the  weights  and  press  key 

TARE 

to  return  to  normal  weighing  

Setting Mode & Description of Parameter Values 

for 3 seconds

,  

  

display momentarily before the display 

reverts to available setting items .There are 12 available setting items (p00-p11) ,indicated

Auxiliary display in the Lower Left corner .  

 

to step through all the available setting items ,and release on

Vetek Weighing AB 

 

weighing  pan  (E.g.  for 

appears, put weights of full load on the weighing pan (E.g. for JWL-6K,  full 

 

mode. 

while powering on the scale. The screen displays 

”, then 

most  digit  blinking),  press  key 

UNIT/LO 

to 

to increase setting values and 

N/G

, the calibration 

” displayed   on 

to  return  to  normal  weighing  

display momentarily before the  display 

p11) ,indicated by 

to step through all the available setting items ,and release on  the 

Summary of Contents for 1.5K

Page 1: ...www cangiaphat com ...

Page 2: ... 2 Display 4 3 Keyboard 4 4 Power Supply 2 3 4 4 5 Operation 5 1 Weighing 5 2 Manual Tare Preset Tare 5 3 Checkweighing 5 4 Simple Counting 5 5 Accumulation Accumulation Display and Accumulation Clearing 5 6 Unit setting menu 4 5 5 6 6 6 6 Calibration 6 1 Single Point Calibration 6 2 Linear Calibration 7 7 7 Setting Mode Description of Parameter Values 8 Serial Interface 8 1 RS 232 Connector 8 2 S...

Page 3: ...ould be performed by authorized personnel only 3 Before Using the Product 3 1 Unpacking and Checking Open the package and check the instrument for transport damage Immediately inform your dealer if you have complaints or if parts are missing The package should contain Scale body Weighing platform plastic base with stainless steel pan Power cord User manual Installing Components 1 Before using the ...

Page 4: ...f low battery and charging status indication Multiple functions tare preset tare simple counting checkweighing and Accumulation Single point calibration and linear calibration available CE type approved Display Model 1 5K 3K 6K 7 5K 15K 30K Capacity Kg 1 5 3 6 7 5 15 30 Division 1 g 0 5 1 2 2 5 10 Division 2 g 0 2 0 5 1 1 2 5 Division 3 g 0 1 0 2 0 5 0 5 1 2 Division 4 g 0 05 0 1 0 2 0 2 0 5 1 Dis...

Page 5: ...Tare or Preset Tare Indication Center of Zero Indication The zeroing range is 2 of weighing capacity Auxiliary display parameter accumulated number of weighments TARE when manual Tare action is done Symbol points at Preset Tare when preset tare value is set Gross weight minus Tare Symbol points at NET or preset are actions are done Under simple counting mode Symbol points at UW when calculated uni...

Page 6: ...d tare entrie M key Adds the indicated weight into Accumulation memory During editing selects the next higher option or increases setting values M key Deletes accumulation records During editing selects the next lower option or decreases setting values ZERO ESC key Zeros the display within 2 of max capacity Exits from setting mode Power Supply Please verify the local AC power source and switch the...

Page 7: ...that must be held in a container tare stores the container weight into memory Manual Tare 1 Under the weighing mode place the container on the weighing pan wait tillstable symbol appears and press the key TARE The container is tared 2 Place the item s to be weighed into the container The weight displayed is the net weight 3 Remove all items from the weighing pan the screen displays the tare value ...

Page 8: ...ainer to display gross and net weight by turns Use this mode to compare the weight of an item to Lower and Upper limits When the checkweigh mode is enabled the indicator will turn on Began by pressing down key UNIT LO for 3 seconds The scale is now in most digit to increase setting values key M to decrease setting value to shift leftwards and key PRINT HI toshift save the Lower limit and return to...

Page 9: ...samples and put the load on the scale begins to count The larger the sample size the more accurate unit weight Symbol points at UW when calculated unit weight is lower than 4 5 of scale Accumulation Accumulation Display and AccumulationClearing Under the weighing mode put the item on the weighing pan Press key is displayed momentarily before the display reverts Remove the item and the display goes...

Page 10: ...t is turned on Press key M to set whether to retain the last set weight unit when the unit is turned on then press key G N to save and return to normal weighing mode Last used weight unit is retained when the unit is turned on Last used weight unit is not retained when the unit is turned on Note if enter the Unit setting menu by method 2 press key G N to save and then press key ZERO ESC to return ...

Page 11: ...d on the weighing pan E g for JWL The calibration procedure ends up with a symbol of Remove all the weights and then press key TARE to return to weighing mode while powering on the scale The screen displays to enter the zero point calibration appears with the right most digit blinking press key to shift rightwards key M to increase setting values and value Put the corresponding weights on the weig...

Page 12: ... are Aut 10 Aut 30 Aut 60 and Aut 90 minutes P01 Set Check Weighing buzzer beep Options are B UP B in B out and b There will be a warning sound when the material s weight exceeds the preset upper There will be a warning sound when the material s weight is between the upper and There will be a warning sound when the material s weight exceeds the preset upper and lower limits and the exceeding weigh...

Page 13: ... 1 division Zero 1 2divisions Zero 2 3divisions Zero 3 4 divisons Zero 4 5divisions and Zero 5 6divisions 11 P10 Initial Setup not available 12 P11 Unit setting refer to section 5 6 8 Serial Interface If external interface is needed please select the proper three in one board first which integrates RTC time display RS 232 and relay weight checking functional module onto one circuit board Only afte...

Page 14: ...contact Service E05 out ran Overload condition Remove weight that is greater than the scale capacity from the pan E07 E Lo Hi The weight checking upper limit value is less than the lower limit value Reset the upper limit value E08 CK UNIT The unit set for weight checking is different from the current unit Disable weight checking function or reset HI LO value EU nit improper accumulation unit Retur...