Vetek Weighing AB
Press key
M+
or key
M-
to choose setting item p11,(There are 12 available setting items
---p00-p11, indicated by Auxiliary display at the Lower Left corner.) then press key
UNIT/LO
to start the unit setting menu.
2) How to set the init weight unit?
Press key
UNIT/LO
to set init weight unit and press key
G/N
to save and move to the next
setting menu---set the active weight unit.
4) How to set the active weight units?
Press key
UNIT/LO
to step through all the available weight units, press key
M+
switch
ON/OFF the displayed weight unit, and press key
G/N
to save and move to the next setting
menu--- whether to retain the last used weight unit when the unit is turned on.
-- enable
-- disable
5) Whether to retain the last used weight unit when the unit is turned on?
Press key
M+
to set whether to retain the last set weight unit when the unit is turned on, then
press key
G/N
to save and return to normal weighing mode.
-- Last used weight unit is retained when the unit is turned on.
-- Last used weight unit is not retained when the unit is turned on.
Note: if enter the Unit setting menu by method 2, press key
G/N
to save and then press key
ZERO/ESC
to return to normal weighing mode.
6. Calibration
1. For best results calibrate the scale at regular intervals. This is especially important if the
scale is in use for prolonged periods.
2. Calibration weights are not provided with the scale.
3. Here we take JWL-15K as an example.
Linear Calibration
1) Press and hold key
TARE
while powering on the scale. The screen displays
”, then
release the
2) Again press key
TARE
to enter the zero point calibration modes, with no load on the
scale.
3) Wait till
appears, put weights of 1/3 of full load on the weighing pan (E.g. for
Summary of Contents for 1.5K