19
Name and Function of Each Part
4
POWER
LED
(green light)
• Lights up when the power is on (p. 31). (Normal operation)
• Rapidly blinks when a checksum error has occurred (p. 95).
5
OVERLOAD
LED
(red light)
• Flashes three times when demagnetizing or automatic zero-
adjustment cannot be performed (p. 91).
• Rapidly blinks when an overload state is detected (p. 90),
the device has entered protection mode (p. 62), or a
checksum error has occurred
(p. 95).
6
JAW UNLOCKED
LED
(red light)
• Lights up when the upper jaw is unlocked.
• Flashes three times when demagnetizing or automatic zero-
adjustment cannot be perfo
rmed (p. 91).
• Rapidly blinks when the device has entered protection mode
(p. 62) or a checksum error has occurred (p. 95).
7
DEMAG / AUTO ZERO
LED
(orange light)
• Slowly blinks when demagnetization or automatic zero-
adjustment has not yet been performed
(p. 33).
• Lights up when demagnetization and automatic zero-
adjustment are in execution
(p. 39).
• Goes out when demagnetization and automatic zero-
adjustment have been completed (p. 39).
• Flashes three times when demagnetization and automatic
zero-adjustmen
t cannot be performed (p. 91).
• Rapidly blinks when the device has entered protection mode
(p. 62) or a checksum error has occurred (p. 95).
8
RANGE
LEDs
(green lights)
• Lights up when its current range is chosen (p. 53). (Normal
operation)
• Blinks rapidly when the device has entered protection mode
(p. 62) or a checksum error has occurred (p. 95).