ALARM MESSAGES
ALARM MESSAGES
Alarm messages are displayed in reverse video.
Press
ALERT RESET
to reset all alarms except LEAD OFF.
The following describes each of the alarm indications.
All audible alarms have been turned off.
LEAD OFF
A lead has been disconnected or the electrode offset potential has exceeded
±
0.5 V.
This
alarm
cannot
be reset with the ALERT RESET key.
HR HIGH
The high heart rate alarm limit has been exceeded for four seconds.
HR LOW
The low heart rate alarm limit has been exceeded for four seconds.
ASYSTOLE
The interval between heartbeats has exceeded six seconds.
22
Model 101R/NR Service Manual