MONITOR SETUP
MONITOR SETUP
To setup the instrument for operation:
1.
Plug the ac line cord into a power source providing the proper voltage.
Check the monitor's rear panel to verify the voltage range required by the monitor.
2.
Press the
POWER
switch at the left side of the front panel to turn power on.
3.
Connect the patient cable to the ECG connector on the front panel.
4.
Connect the Trigger Output from the monitor rear panel to the device being triggered.
To change Mains Voltage:
1.
Locate power entry module in Rear Panel.
2.
Remove line cord.
3.
Pry down entry cover.
4.
Locate and remove voltage wheel and set for appropriate voltage.
5. Replace fuse with proper size (see fuse rating).
6. Replace
cover.
12
Model 101R/NR Service Manual