ECG MONITORING
×
ECG
SIZE
MM/MV
Ø
LEAD
II
FILT
3.
Use the top two programmable keys to the right of the display to adjust the ECG waveform amplitude.
×
Increases ECG size
Ø
Decreases ECG size
Lead Selection
Select [
LEAD
] to change the lead selection. The current lead selection is shown below the alarm limits in the upper
right portion of the display.
Available lead selections are Lead I, Lead II, or Lead III.
ECG Filter
Select [
FILTER
] to turn the filter on or off. The filter on/off indication is shown in the upper right portion of the
display. The filter sets the frequency response of the displayed waveform as follows:
Filtered:
0.5 to 25 Hz
Unfiltered:
0.5 to 45 Hz
Model 101R/NR Service Manual
16
Summary of Contents for 101NR
Page 1: ...Service Manual Model 101R NR Patient Monitor R wave Trigger 2409 00 20 ...
Page 15: ...MONITOR DESCRIPTION MENU STRUCTURE HIGH LOW TRIG DELAY 101R Model 101R NR Service Manual 9 ...
Page 56: ...BOARD LAYOUT DIAGRAMS Model 101R NR Service Manual This page is intentionally left blank ...
Page 57: ...BOARD LAYOUT DIAGRAMS Model 101R NR Service Manual A BOARD LAYOUT DIAGRAMS ...
Page 58: ......
Page 59: ......
Page 60: ......
Page 61: ...SCHEMATICS Model 101R NR Service Manual B SCHEMATICS ...
Page 62: ...SCHEMATICS Model 101R NR Service Manual This page is intentionally left blank ...
Page 63: ......
Page 64: ......