2. Safety First

This chapter is especially important for
personnel who work with the 
SR Ivocap System or who have to carry out
maintenance or repair work. This chapter
must be read and the corresponding 
instructions followed.

2.1 Indications

The SR Ivocap System must only be used for
injecting the SR Ivocap material and it should
be used for this purpose only. Other uses
than the ones stipulated, i.e. injecting other
materials, etc. are contraindicated. The
manufacturer does not assume any liability
for damage resulting from misuse. The user
is solely responsible for any risk resulting
from failure to observe these Instructions. 

Further instructions to assure
proper use of the system:
– The instructions, 

regulations, and notes in
these Operating 
Instructions must be 
observed.

– The system must be 

operated under the 
indicated environmental
and operating conditions
(Chapter 9).

– The SR Ivocap System

must be properly 
maintained (Chapter 7).

2.1.1

Risk of crushing

Do not reach between
pressing plunger and safety
loop with your fingers. There
is a risk of crushing when air
is compressed.

Use protective thermo-gloves
to touch hot metal parts. 

Do not use the water bath
without the plastic floaters.
Splash water presents a burn
hazard.

Wear protective clothing to
protect yourself from splash
water.

2.1.2

Burn hazard

2.1.3

Burn hazard

2.1.4

Burn hazard

Summary of Contents for SR Ivocap

Page 1: ...Operating Instructions SR Ivocap Equipment System...

Page 2: ...2...

Page 3: ...lation 8 4 1 Unpacking and checking the contents 4 2 Selecting the location 5 Operation 9 5 1 Operating elements at the pressure apparatus 5 2 Operating elements at the clamping frame 6 Procedure 10 6...

Page 4: ...eter A3 Extrusion piston A4 Vise shank A5 Safety loop B Clamping frame JS3 B2 Thrust collar with lever B3 Pressing plate C Flask C1 Flask half M C2 Flask half W C3 Flask lid C4 Injection funnel D Caps...

Page 5: ...rmation Contraindication 1 3 Notes regarding the Operating Instructions These Operating Instructions must be read prior to working with the SR Ivocap System Equipment concerned SR Ivocap System Target...

Page 6: ...y risk resulting from failure to observe these Instructions Further instructions to assure proper use of the system The instructions regulations and notes in these Operating Instructions must be obser...

Page 7: ...as been especially developed to compensate curing shrinkage Controlled polymerization allows shrinkage of the material during polymerization to be compensated by inflowing material The SR Ivocap injec...

Page 8: ...poses 4 2 Selecting the location The diagram below depicts one possible installation Important information For safety reasons the installation of the pressure power and water connections may only be c...

Page 9: ...sed horizontal position 5 2 Operating elements on the clamping frame Thrust collar with lever B2 Operation 1 Thrust collar with lever closed push to the right until it stops 2 Thrust collar with lever...

Page 10: ...oling in a cold water bath at least 30 minutes the injection pressure must be maintained for at least 20 minutes thereof 3 2 1 Compressed air 6 bar Water level Pressure apparatus Capsule Flask Mark fo...

Page 11: ...air connection before maintenance and cleaning 7 1 Monitoring and maintenance The time for these maintenance procedures depends on the frequency of use and work ing habits of the users For that reaso...

Page 12: ...tes for users Flask margins are not clean plaster residue on the injection funnel The model should be trimmed prior to investment Clamping frame is not correctly positioned i e in the centre of the pr...

Page 13: ...Acceptable operating conditions Acceptable humidity range Maximum relative humidity at 31 C 87 F gradually decreasing to 50 at 40 C 104 F condensation excluded Acceptable ambient pressure The unit is...

Page 14: ...14...

Page 15: ...15...

Page 16: ...627 33 99 Fax 57 1 633 16 63 www ivoclarvivadent com Ivoclar Vivadent SAS B P 118 F 74410 Saint Jorioz France Tel 33 450 88 64 00 Fax 33 450 68 91 52 www ivoclarvivadent fr Ivoclar Vivadent GmbH Dr Ad...

Reviews: