Ivoclar Vivadent SR Ivocap Operating Instructions Manual Download Page 12

12

8. What if...?

8.1 Error messages 

List of possible error messages and their meaning:

This chapter will help you to recognize
malfunctions and take appropriate measures
or, if possible, to perform some repairs.

8.2 Repairs

Repairs may only be carried out by a certified Ivoclar Vivadent Service Center. 
Please refer to the addresses in Chapter 10.

If repairs during the warranty period are not carried out by a certified 
Ivoclar Vivadent Service Center, the warranty will expire immediately.

Please also read the safety information in Chapter 2.

Error

1

2

3

4

5

6

7

8

9

10

11

Flasks do not close.
The flask covers do not sit on the flask margins.

The model to be invested does not fit in the flask.

The pressure is difficult to be transferred to the flask and/or clamping frame.

Pressure in the clamping frame fluctuates.

Injection plunger enters the capsule and deforms the bottom of the capsule. The material is
extruded at the wrong end of the capsule.

The rod of the injection plunger cannot be moved, even if the connection to the compressed
air source is interrupted. It seems to be stuck.

Injection plunger does not move. The SR Ivocap material is not injected.

The resin flows out of the flask during polymerization.

The locking mechanism is blocked.

The locking mechanism is contaminated with wax.

Capsule reservoir has polymerized after 35 minutes.

Description

Notes for users

Flask margins are not clean; plaster residue on the injection funnel.

The model should be trimmed prior to investment.

Clamping frame is not correctly positioned, i.e. in the centre of the
press.

Compression spring in the clamping frame are old or dirty and cannot
transfer the pressure.

Replace the compression spring or clean the

clamping frame (see Error 12).

The injection plunger is not correctly placed on the capsule. There is 
an empty space between the injection plunger and capsule bottom.
Avoid empty space and place the injection plunger evenly on the 
capsule bottom.

– No compressed air present
– Capsule cover was not removed
– There is plaster residue or the funnel is blocked, e.g. by plaster 

residue

– The injection channels are too narrow.
– The flow properties of the resin are poor because of incorrect 

storage

– Leaks as a result of incorrect investment.
– Flask defective.
– Inadequate clamping pressure on the frame.
– A foreign object is stuck between the two flask halves.

The mechanical components are damaged as a result of improper 
handling. In case of considerable damage, the mechanical components
must be replaced.

Wax residue in the water bath has contaminated the clamping frame.
Boil out clamping frame with clean, hot water.

– Water level surpasses the marking on the clamping frame.
– An unsuitable polymerization bath was used.
– The system was not placed in cold water immediately after 

polymerization.

– Reduce polymerization time by 1–2 minutes if necessary

Negative pressure in the pressure cylinder may prevent the plunger
from moving, open valve

Summary of Contents for SR Ivocap

Page 1: ...Operating Instructions SR Ivocap Equipment System...

Page 2: ...2...

Page 3: ...lation 8 4 1 Unpacking and checking the contents 4 2 Selecting the location 5 Operation 9 5 1 Operating elements at the pressure apparatus 5 2 Operating elements at the clamping frame 6 Procedure 10 6...

Page 4: ...eter A3 Extrusion piston A4 Vise shank A5 Safety loop B Clamping frame JS3 B2 Thrust collar with lever B3 Pressing plate C Flask C1 Flask half M C2 Flask half W C3 Flask lid C4 Injection funnel D Caps...

Page 5: ...rmation Contraindication 1 3 Notes regarding the Operating Instructions These Operating Instructions must be read prior to working with the SR Ivocap System Equipment concerned SR Ivocap System Target...

Page 6: ...y risk resulting from failure to observe these Instructions Further instructions to assure proper use of the system The instructions regulations and notes in these Operating Instructions must be obser...

Page 7: ...as been especially developed to compensate curing shrinkage Controlled polymerization allows shrinkage of the material during polymerization to be compensated by inflowing material The SR Ivocap injec...

Page 8: ...poses 4 2 Selecting the location The diagram below depicts one possible installation Important information For safety reasons the installation of the pressure power and water connections may only be c...

Page 9: ...sed horizontal position 5 2 Operating elements on the clamping frame Thrust collar with lever B2 Operation 1 Thrust collar with lever closed push to the right until it stops 2 Thrust collar with lever...

Page 10: ...oling in a cold water bath at least 30 minutes the injection pressure must be maintained for at least 20 minutes thereof 3 2 1 Compressed air 6 bar Water level Pressure apparatus Capsule Flask Mark fo...

Page 11: ...air connection before maintenance and cleaning 7 1 Monitoring and maintenance The time for these maintenance procedures depends on the frequency of use and work ing habits of the users For that reaso...

Page 12: ...tes for users Flask margins are not clean plaster residue on the injection funnel The model should be trimmed prior to investment Clamping frame is not correctly positioned i e in the centre of the pr...

Page 13: ...Acceptable operating conditions Acceptable humidity range Maximum relative humidity at 31 C 87 F gradually decreasing to 50 at 40 C 104 F condensation excluded Acceptable ambient pressure The unit is...

Page 14: ...14...

Page 15: ...15...

Page 16: ...627 33 99 Fax 57 1 633 16 63 www ivoclarvivadent com Ivoclar Vivadent SAS B P 118 F 74410 Saint Jorioz France Tel 33 450 88 64 00 Fax 33 450 68 91 52 www ivoclarvivadent fr Ivoclar Vivadent GmbH Dr Ad...

Reviews: