Ivoclar Vivadent SR Ivocap Operating Instructions Manual Download Page 1

Operating Instructions

SR Ivocap

®

Equipment System

Summary of Contents for SR Ivocap

Page 1: ...Operating Instructions SR Ivocap Equipment System...

Page 2: ...2...

Page 3: ...lation 8 4 1 Unpacking and checking the contents 4 2 Selecting the location 5 Operation 9 5 1 Operating elements at the pressure apparatus 5 2 Operating elements at the clamping frame 6 Procedure 10 6...

Page 4: ...eter A3 Extrusion piston A4 Vise shank A5 Safety loop B Clamping frame JS3 B2 Thrust collar with lever B3 Pressing plate C Flask C1 Flask half M C2 Flask half W C3 Flask lid C4 Injection funnel D Caps...

Page 5: ...rmation Contraindication 1 3 Notes regarding the Operating Instructions These Operating Instructions must be read prior to working with the SR Ivocap System Equipment concerned SR Ivocap System Target...

Page 6: ...y risk resulting from failure to observe these Instructions Further instructions to assure proper use of the system The instructions regulations and notes in these Operating Instructions must be obser...

Page 7: ...as been especially developed to compensate curing shrinkage Controlled polymerization allows shrinkage of the material during polymerization to be compensated by inflowing material The SR Ivocap injec...

Page 8: ...poses 4 2 Selecting the location The diagram below depicts one possible installation Important information For safety reasons the installation of the pressure power and water connections may only be c...

Page 9: ...sed horizontal position 5 2 Operating elements on the clamping frame Thrust collar with lever B2 Operation 1 Thrust collar with lever closed push to the right until it stops 2 Thrust collar with lever...

Page 10: ...oling in a cold water bath at least 30 minutes the injection pressure must be maintained for at least 20 minutes thereof 3 2 1 Compressed air 6 bar Water level Pressure apparatus Capsule Flask Mark fo...

Page 11: ...air connection before maintenance and cleaning 7 1 Monitoring and maintenance The time for these maintenance procedures depends on the frequency of use and work ing habits of the users For that reaso...

Page 12: ...tes for users Flask margins are not clean plaster residue on the injection funnel The model should be trimmed prior to investment Clamping frame is not correctly positioned i e in the centre of the pr...

Page 13: ...Acceptable operating conditions Acceptable humidity range Maximum relative humidity at 31 C 87 F gradually decreasing to 50 at 40 C 104 F condensation excluded Acceptable ambient pressure The unit is...

Page 14: ...14...

Page 15: ...15...

Page 16: ...627 33 99 Fax 57 1 633 16 63 www ivoclarvivadent com Ivoclar Vivadent SAS B P 118 F 74410 Saint Jorioz France Tel 33 450 88 64 00 Fax 33 450 68 91 52 www ivoclarvivadent fr Ivoclar Vivadent GmbH Dr Ad...