53
5. Operation and Configuration
2. Conduct Software update
If the USB stick with the software file is already connected, the
furnace automatically searches for a valid software file. If the USB
stick has not yet been connected with the furnace, do so now.
Press the
[Execute]
button.
3. The status bar shows the progress of the update
4. End Software update
The following messages are displayed:
Software update successful
Software update failed
The furnace has to be switched off and on again by means of the mains switch at the rear of the furnace to
complete the software update.
Important information
Please note that modified Ivoclar Vivadent programs may be overwritten during a software update. Individual
programs are not affected and will not be overwritten.
5.4.5 Software update
Software updates can be easily installed on the furnace by means of a USB stick. A USB stick, which contains a current
software file (e.g. P710 G2_V1.10.iv) is required. The software version on the USB stick must be higher than the one
installed on the furnace (see Selection – Information)
The free software updates for Programat furnaces are available from
www.ivoclarvivadent.com/downloadcenter
.
Back-up your data before performing a software update.
A software update requires only few steps:
1. Open Software update
Scroll to page 4 in the home screen and press the
[Software update]
button.