will be returned first. when you read all the error tips information in the error
queue. The ERROR led will turn off. When error occur the buzzer of the
power supply will beep once.
⚫
If more than 20 errors occurred, the last one stored in the queue will be
replaced by “-350”,
Meaning "too many mistakes.". If the error
messages in the error message queue are not read out, the error
following will not be saved to the queue. If there is no error message,
when read error information, it will r
eturn “+0”, meaning “no error”.
⚫
If turn off the power supply or send “CLS(clear status) order, the error
messages in the error queue will be cleared. *RST order will not clear the
error messages in the error queue.
SYSTem:CLEar
This command is used to clear the error information.
Command syntax:
SYSTem:CLEar
Parameters
None
Returns:
None
SYSTem:REMote
This command is used to set the power to remote control mode. Only the Shift,
Local, Select, and On / Off keys are available on the front panel, and all other
keys are locked for use. Failure to send this command for remote control
configuration before sending control commands using a PC may cause
communication errors.
Command syntax
SYST:REM
Parameters
None
Query syntax
None
SYSTem:LOCal
This command is used to switch the instrument to local control