25
12. Power supply scope test
◆
Test Requirements: input AC ~100-240V, measure DC voltage.
◆
Test Standard: debug sample machine to DC working voltage normal, function well.
13. High voltage debug
Power source L/N line add 1500V/5mA AC to floor, continue 60s, high voltage tester doesn’t
warn, the tested equipment no damage, no creepage.
14. Low impedance test
Use low impedance tester to test earthing resistance, the value should less than 0.1Ω.