background image

11

12

          
Power-Up Options

Button

Function

Disables automatic power-off

function Display shows PoFF until the button is released

SELECT

Disable

Beep

function

ENTER

Function Description

Holding a button down while turning the Meter on activates a power-up option. Power-up options allow you to use additional 
features and functions of the Meter. To select a power-up option, hold down the appropriate button indicated while turning the 
Meter from OFF to any switch position. Power-up options are cancelled when the Meter is turned Off.

          
Automatic Power Off

The Meter have automatic power off function (Sleep mode) to conserve battery power. if there is no function change or button 
press for 10 minutes. The Meter comes out of Sleep mode when a key is pressed or when the rotary switched is changed.
To disable the Sleep mode, hold down SELECT button while turning the Meter on. Sleep mode is always disabled in the 
insulation test mode, or if the auto power off feature has been disabled by pressing SELECT button when the Meter is turned 
on.

          
Hold Function

Press Hold button to freeze the displayed value. Press again to release the display.

          
Relative Measurement

Show the difference between actual value and the relative base. 
Press REL Button to enter relative measurement and the meter will record the initial value when pressing the key. 
Displayed value = Actual value -Initial value
Press REL Button again to exit relative measurement.
Press SELECT button to view Initial value, when the Initial value is invalid, display----.

Insulation Test Lock

In insulation test mode, press Test button to perform insulation test until the button is released. when the button is released, 
the screen display hold sign. 
Press Lock Button, then the screen display Lock sign, press Test Button, the meter will perform insulation test unitl you press 
Test button again; The test lock will unlocked while to cancel insulation test.
Lock function is invalid when timer function is selected.

          
Storing Test Data

Press

MEMO

button, the

meter

enter

Hold

status

automatically

.

LCD

screen

display ‘MEMO’ and

the

memory

code

on

secondary

display (see

below). Press

△/▽

button

to

change

the

code, press

enter

to

save

the

data

in

the

preferred

memory

code

and

the

screen

display ‘SAVE’

.

The

data

is

stroed

successfully

when

the

buzz

beeps

.

The

meter

can

save

20 datas

and

memory

code

from

00

to

19.

          
Reading Test Data

Press

READ

button

to

display

the

data

in

memory

.

Press

△/▽

button

to

change

the

code

and

read

the

data

accordingly.

Summary of Contents for Mastech MS5203

Page 1: ...Digital Insulation Tester MS5203 CA T IV 600 V CA T I I I 1000 V HOLD HOLD LOCK DIGITAL INSULATION TESTER MS5203...

Page 2: ...Off Hold Function Relative Measurement Insulation Test Lock Storing Test Data Reading Test Data Delete Data Compare Function Timer Function MAX MIN AVG DAR PI 03 04 05 12 03 04 05 06 08 09 10 10 11 11...

Page 3: ...h ground Use caution with voltages above 30 V ac rms 42 V ac peak or 60 V dc These voltages pose a shock hazard Replace the battery as soon as the low battery indicator appears Disconnect circuit powe...

Page 4: ...soon as the low battery indicator appears Indicates a test lock will be applied the next time you press Test Button on the meter the test lock acts to hold down the button until you press Test Button...

Page 5: ...nals Continuity test function is selected Auto power off function is selected TEST Timer unit min sec Measure unit Display the Max Min Avg value MAX MIN AVG GM V Indicator Description Appears on prima...

Page 6: ...DAR PI Active the timer function Active hold function in DCV ACV Continuity or lock function Insulation Active rel function in dcv acv continuity mode change cursour position Select one test output vo...

Page 7: ...Relative Measurement Show the difference between actual value and the relative base Press REL Button to enter relative measurement and the meter will record the initial value when pressing the key Dis...

Page 8: ...limited value press ENTER to adjust the value The meter will be flashing with the upper limited value or the lower limited value Switch to the proper range or signal polar by pressing SELECT Press to...

Page 9: ...fter applying the test voltage R15Sec Resistance value measured 10 minutes after applying the test voltage After performed insulation test press SELECT button to view DAR or PI If DAR or PI Value inva...

Page 10: ...ce To avoid possible damage to the Meter or to the equipment under test disconnect circuit power and discharge all high voltage capacitors before testing for continuity Switch rotary switch to positio...

Page 11: ...on if meter display please replease battery Insert test leads to High Low terminals if the meter display Live indicator the meter cannot measure on live circuit please power off Live circuit Press TES...

Page 12: ...sulation Source Voltage 20 0 Insulation Short Circuit Test Current 1 8mA nominal Insulation Auto Discharge Discharge time 2 second for C 1 F or less Insulation Live Circuit Detection Inhibit test if t...

Page 13: ...20 3 5 0 20M 20 100M 0 01M 0 1M 1mA 100K 250V 0 20 3 5 0 20M 20 200M 0 01M 0 1M 1mA 250K 200 250M 1M 500V 0 20 3 5 0 20M 20 200M 0 01M 0 1M 1mA 500K 200 500M 1M 1000V 0 20 3 5 0 200M 1 00 5 00G 0 1M 0...

Reviews: