Insulation Test
To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as
appears.
Turn the rotary switch to OFF and remove the test leads from the terminals.
Replacing Battery
1
2
Screw
Battery
Insulation tests should only be performed on dead circuits. Check the test leads before testing.
Switch rotary to Insulation position. if meter display
, please replease battery.
Insert test leads to High
、
Low
terminals
.
if
the
meter
display
Live,
indicator
the
meter
cannot
measure
on
live
circuit.
please
power
off
Live
circuit.
Press TEST.V button to select output voltage.
In insulation measuring,
the screen alternate turn on or off, the primary screen display the resistance value, and the
secondary screen display the Output voltage. Release the test button then discharges through the Meter, the meter display
DISC.
The secondary Display 0 VDC indicator when the voltage discharge finished.
Disconnect the test leads from circuit.