Page 20 of 43
Low Temp Furnace Model ITL-M-17701 13-07/20
The NETWORK INTERFACE:
Access to the NETWORK interface is available when logged in as the ENGINEER. This will allow
adjustment to various network parameters, should they need to be adjusted.
This menu allows writeable access to:
•
IP type (FIXED or DYMAMIC)
•
IP address
•
Subnet Mask
•
Gateway
and Read only access to:
•
MAC address
•
Client identifier
To access the NETWORK INTERFACE:
1. Log in as the ENGINEER
2. From the ENGINEER page scroll to NETWORK tab and enter with the SCROLL button
3. Scroll to INTERFACE and enter with the scroll button
4. Scroll to and adjust desired parameter using the pop out soft keyboard
DATA LOGGING:
The equipment will always record data for all four channels UNLESS the equipment is in ENGINEERING
mode. This is indicated by the green
R
in the bottom left had side of the ISOTECH homepage (Figure 7)
The data can either be extracted via the USB socket on the front panel or to a PC by means of the FTP
protocol (Remote Archiving) using REVIEW software, see page 42 for details.
To back up the data via USB stick on the control panel use the following procedure:
1. Log in as the SUPERVISOR
2. Plug a suitable USB memory stick to the socket on the control panel (Figure 1)
3. Ensure the controller has accepted the memory stick by the icon in the lower left of the ISOTECH
homepage
4. Press the page button to scroll to the SUPERVISOR menu
5. Scroll to DEMAND ARCHIVING
6. Scroll to ARCHIVE TO and enter, select USB from the options and enter
7. Scroll to ARCHIVE and select from:
a)
Bring to date
b)
All
c)
Last month
d)
Last week
e)
Last day
f)
Last hour
g)
None
8. When the selection is entered by pressing the SCROLL button, the transfer will begin (this may take
several minutes depending on the selection made)
9. When the transfer is complete, remove the USB stick and exit SUPERVISOR mode
The data will be stored in a folder called HISTORY.