BATTERY REPLACEMENT

 

    Figure 2

 

Case Bottom 

Case Top 

Battery Connector 

 1.5V x 2 Battery 

 Battery Box 

Summary of Contents for IDM63N

Page 1: ...INSTRUCTION MANUAL IDM63N DIGITAL MULTIMETER EN IT DE FR...

Page 2: ...1 ISO TECH IDM 63N DIGITAL MULTIMETER INSTRUCTION MANUAL...

Page 3: ...2...

Page 4: ...items 1 Digital Multimeter 2 Test lead one black one red 3 Instruction Manual 1 2 Meter Safety Terms as Marked on Equipment ATTENTION Refer to Manual DOUBLE INSULATION Protection Class II DANGER Risk...

Page 5: ...nput Terminal Positive input connector for Volts Ohms and Diode and currents 5 Range Switch Manual Range The RANGE switch is pressed to select manual ranging and to change ranges When the RANGE switch...

Page 6: ...5 Figure 1...

Page 7: ...nly be guaranteed while the limits of section 2 2 are observed Display The Liquid Crystal Display LCD has a maximum reading of 3200 and 65 segments bar graph Polarity Indication Automatic positive imp...

Page 8: ...R H 30 C 40 C 75 R H 40 C 50 C 45 R H Storage Temperature 20 C to 60 C 0 to 80 R H with battery removed from the meter Temperature Coefficient 0 15 x Specified accuracy C 18 C or 28 C Power Requiremen...

Page 9: ...ccuracy is reading number of digits at 23 C 5 C less than 80 R H 1 DC Volts Range Resolution Accuracy Over voltage protection 300mV 100 V 0 5 reading 2digits 3V 1mV 30V 10mV 300V 100mV 600V 1V 600V DC...

Page 10: ...5 reading 5digits 600V DC or 600V rms 30V 1 5 reading 5digits 40Hz to 500Hz 300V 600V Resolution 1mV 10mV 100mV 1V Frequency Response 40Hz 300Hz for 3V Range Input Impedance 10M less than 100PF The r...

Page 11: ...stance Range Resolution Accuracy Overload Protection 300 0 1 1 0 reading 4digits 600V DC or 600V rms 3K 1 30K 10 300K 100 3M 1K 30M 10K 2 0 reading 5digits 0 8 reading 2digits Open Circuit Voltage 1 3...

Page 12: ...ge Resolution Accuracy Max Test Current Max Open Circuit Voltage 1mV 1 5 reading 5digits 1 5mA 3 3V For 0 4V 0 8V Overload Protection 600V DC AC rms max Continuity Internal sounder operates when resis...

Page 13: ...erload Protection 300 A 0 1 A 1 0 reading 2digits 4 5mV A 25 C 3000 A 1 A 600V rms or 3200 A rms 6 Auto Power Off The meter will automatically shut off approximately 10 minutes after power on The mete...

Page 14: ...to retain the instrument in safe condition 3 1 Preparation and Caution before Measurement 1 Allow at least 60 seconds after switching on before taking measurements 2 When the function switch selector...

Page 15: ...he use of fused test leads constructed in accordance with the HSE guidance note GS38 Electrical Test Equipment for use by Electricians should be used 3 2 Voltage Measurements 1 Connect the red test le...

Page 16: ...e that the device being tested has no voltage present 4 Connect the test leads across the resistor to be measured To ensure the best accuracy in measurement of low resistance short the test leads befo...

Page 17: ...ween 400V to 900V If the diode under test is defective 000 short circuit or OL non conductance is displayed Reverse Check of Diode If the diode under test is good 1 is displayed If the diode under tes...

Page 18: ...aware of the hazard involved Whenever it is likely that the protection has been impaired the instrument must be made inoperative and be secured against any unintended operation The protection is like...

Page 19: ...eter off Remove the test leads from the input terminals 2 Position the meter face down Remove the screw from the case bottom 3 Lift the end of the case bottom until it gently unsnaps from the case top...

Page 20: ...19 BATTERY REPLACEMENT Figure 2 Case Bottom Case Top Battery Connector 1 5V x 2 Battery Battery Box...

Page 21: ...20 HOW TO USE THE PROBE HOLDER Slide out one probe holder for one handed meter operation...

Page 22: ...21 HOW TO USE THE PROBE HOLDER Wrap the leads around the holster to store the test probes...

Page 23: ...22 HOW TO USE THE TILT STAND AND HOLSTER Hang on a nail at the workbench Swing the stand out for easier meter reading Swing the upper holder out and hook it over a door...

Page 24: ...nited Kingdom Japan West Tower 12th Floor Yokohama Business Park 134 Godocho Hodogaya Yokohama Kanagawa 240 0005 Japan USA 7410 Pebble Drive Fort Worth Texas 76118 6961 Canada 1701 Woodward Drive Ste...

Reviews: