ISG 601-502 Instructions Manual Download Page 4

Spare bulbs: 3W LED lamp 

Spare fuse: 250V 1A delayed model, 5 X 20 fuse 

8. Mounting the bulb 

Never touch the surface of the bulb with your hand directly. Use gloves or a cloth when you mount the 

bulb to avoid leaving fingerprints. Fingerprints or stains should be wiped off with a tissue moistened 

with  a  small  amount  of  alcohol.  The  fingerprints  will  etch  on  the  bulb's  surface  and  make  the 

brightness lower and life cycle shorter. 

Please  make  sure  the  bulb's  touching  points  are  Ok  as  it  will  not  light  or  short  circuit  if  the  bulb's 

touching points are damaged. Insert the bulb's legs into the bulb holder firmly. The bulb will go out if its 

legs have not been inserted into the holder firmly. Such Operation may cause short circuit or smoking.   

9. Temperature for illumination 

When  you  use  the  halogen  bulb,  there  will  be  high  temperature  when  the  bulb  is  lightened,  never 

move  the  collector  when  the  bulb  is  lightened.  Do  not  touch  the  bulb  within  30  minutes  after  it  has 

been gout out.  Make sure the bulb has  been be cooling enough time (at least  30 minutes) to avoid 

scald. 

Do not put fibre, papers and incendive things (eg. Gasoline, aether, methanol and ethanol) close to 

the bulb. 

10.Make  sure  to  turn  off  the  power  supply  before  assembling 

microscope, replacing bulb or fuse.

 

11.Observing with oil immersion 

Using small amount of oil immersion is enough. The redundant oil will adhere to stage or 
condenser and these will reduce microscope’s performance. Get rid of the redundant oil or clean the 

lens  by  using  aether  or  pure  alcohol  according  to  this  instruction.  Pay  attention  during  process  as 

these things are incentive. The instruction you buy may not including some products mentioned in this 

instruction.  Safety  has  been  considered  during  design  while  users  still  have  the  possibility  to  be 

injured  and  the  instruction  to  be  damaged  when  improper  operation  made.  Please  pay  much  more 

attention to read this instruction carefully before using and keep it properly to make sure it will be got 

when it is necessary. 

 

 

Summary of Contents for 601-502

Page 1: ...optimum performance and familiarize yourself with the use of this microscope we recommend that you read this manual thoroughly before operating the microscope Retain this instruction manual in an easily accessible place near the microscope for further reference ...

Page 2: ...Contents 1 Structure and nomenclature 2 Application 3 Technical parameter and specification 4 Configuration 5 Assemblage 6 Operation 7 Troubleshooting 8 Maintenance ...

Page 3: ...cope with the dust cover 6_Do not place microscope in moist room to avoid short circuit 3 Focusing Dials Never turn the focusing dials in the adverse direction at the same time Do not turn the coarse focusing knob when the stage reaches max position These incorrect operations will make focusing structure unworkable 4 Clamp The clamps fix the microscope firmly during transportation Please remove th...

Page 4: ... lightened Do not touch the bulb within 30 minutes after it has been gout out Make sure the bulb has been be cooling enough time at least 30 minutes to avoid scald Do not put fibre papers and incendive things eg Gasoline aether methanol and ethanol close to the bulb 10 Make sure to turn off the power supply before assembling microscope replacing bulb or fuse 11 Observing with oil immersion Using s...

Page 5: ...g the microscope Unproper operation will cause hurt or damage to the operator and instrument ___ 3 Switch on 4 Switch off 13 Operating the microscope only indoor and under the adtitude 2000m IP52 grads If the equipment is used in manner not specified by the manufacturer the provided by the equipment may be impaired ...

Page 6: ...Objectives 6 Mechanical stage 7 Condenser fixing screw 8 Condenser 9 Collector 10 Observation Head 11 Observation head fixing screw 12 Microscope body 13 Slide clamp 14 Y direction stage movement 15 X direction stage movement 16 Fine focusing knob 17 Coarse focusing knob 18 Iris diaphragm 19 Trinocular tube ...

Page 7: ...ogy observation in medical and hygienic establishments chemical laboratory research institute and college ect This product with modern design good looking appearance adopts many advanced structure technology in order that the clients can operate the instrument conveniently and safely The following accessories can be optional for the customer darkfield phase contrast photographic and video accessor...

Page 8: ...b control range stage will be up or down 0 2mm per circle Coarse focusing knob control range stage will be up or down 36mm 4 Mechanical stage X movement range 75mm Y movement range 50mm 5 Nosepiece Roller bear quintuple nosepiece 6 Condenser Abbe NA 1 25 with iris diaphragm 7 Eye tubes Interpupillary range 55 75mm 8 Working environment Temperature 0 40 Humidity max 85 no dew 2 Optical parameter 1 ...

Page 9: ...ace Resolution λ 2XN A λstands for wavelength λ 0 55um 4 Working distance The distance between the object surface to the first surface of optical system when you get the sharpest image The W D will become more short when the magnification for the objective is bigger 5 Eyepiece view field The max diameter measured when the diaphragm is open 10X 18 means the magnification is 10X and the max diameter...

Page 10: ... diaphragm without auxiliary lens 1 pc 4 Objectives achromatic objectives 4X10X40X100X 1 set 5 WF10X eyepiece 2 pcs 6 Blue filter 1 pc 7 Power cord 1 pc 8 Immersion oil 1 bottle 9 Dust cover 1 pc 10 Instruction 1 pc 11 Wrench 2 5 1 5 1 set Optional Item Content 1 Collector with iris diaphragm 2 Darkfield condenser 3 WF10X eyepiece with reticle 4 WF15X eyepiece 5 Plan objectives 4X 10X 40X 100X 6 6...

Page 11: ...pull it out Remove the clamps by wrenches _ Mechanical stage There is one clamp fixing the mechanical stage in Y axis direction Remove the bolts and clamps _ Focusing adjustment Pull out the collector The bracket is fixed by one clamp Please remove the bolts and clamps 3 Mount the binocular head Loosen the fixing screw in the eye tubes and insert the binocular head into the eye tubes Then tighten ...

Page 12: ...es supplied by manufacturer to remove the clip 4_Eyepieces 10X eyepieces have been in the foam box before shipment Insert the eyepiece into the eyepiece tubes and tighten the fixing screws by wrench 5_Other attachments Replace other attachments eg camera according to the instruction 6 Replacement 1_Replacement for bulb halogen bulb optional Turn off the power supply turn the switch to _O_ and pull...

Page 13: ...ase read the instruction carefully to avoid making damage to the cover glass or objective 1_Put the 10X or 4X objective into optic path 2_Remove the stage to the top by turning the coarse focusing knob 3_Observe through the eyepiece and turn the coarse focusing knob slowly to let the stage down and down till there is sharp image 4_Turn the fine focusing knob to focus finely Use the 10X objective o...

Page 14: ...rture diaphragm adjusting shaft for condenser till the position which stands for the magnification of objective _ adjusting the aperture diaphragm 1_Adjust the aperture diaphragm by the adjusting shaft The brightness and resolution will be decrease while the contrast and depth of view will increase if the diameter of aperture diaphragm reduces 2_Adjust the value of aperture diaphragm to 70 80 of t...

Page 15: ...bserve the exit pupil which is light and in nummular shape c Remove the air bubble as following Turn the nosepiece slowly and turn the oil immersion objective once or twice Turn the lifting knob for condenser slowly and make the condenser a little bit of up and down Increase the amount of oil immersion or change the old oil immersion with new one 3_Operation for oil immersion Use the oil immersion...

Page 16: ...ant decrease the tension please turn the ring with ershins The stage will fall down automatically if the tension of the ring decreased too much VII Trouble Shooting If difficulties should be encountered in the course of operation and no major instrument mal functions can initially be detected please recheck the symptoms referring to the tables provided below before contacting the service departmen...

Page 17: ...own Turn over the slide Immersion oil on dry objective especially 40X Cleaning No immersion oil used on immersion objective Use immersion oil Dirt or dust on the lens condenser objective eyepiece or collector Cleaning Air bubbles in immersion oil Remove air bubbles Unspecified immersion oil used Use specified immersion oil Condenser aperture and field diaphragm opened too far Close properly Dirt o...

Page 18: ...Adjustment Excessive eye fatigue Incorrect diopter adjustment Correct adjustment Inadequate brightness or illumination Adjust brightness with control dial 3 Electrical Symptoms Causes Countermeasures Lamp does not light when switched on No electrical power Check power cord connection Lamp bulb not inserted Insert correctly Lamp failure Replacement Sudden lamp failure Unspecified lamp bulb used or ...

Page 19: ...ng painted or plastic surfaces Avoid use of and organic solvents such as alcohol ether thinner ect to clean the painted or plastic surfaces of the instrument We recommend the use of silicon cloth More persistent dirt may be cleaned with mild detergent solution Printed plastic surfaces should be cleaned only by soft cloth moistened with water 3 When not in use When the microscope is not in use cove...